English
Language : 

MDS-040APS24B Datasheet, PDF (4/10 Pages) Delta Electronics, Inc. – MDS Medical AC-DC Open Frame
T``E`CHNICAL DATASHEET
MDS Medical AC-DC Open Frame
24Volt, 40Watt / MDS-040APS24 B
Protections
Overvoltage (max)
Over load / Over current (max)
Over Temperature
Short Circuit
Protection Against Shock
*PE: Protective Earth
Reliability Data
MTBF (typ.)
Safety Standards / Directives
Medical Safety
ITE Safety
CE
Material and Parts
Galvanic Isolation
150%, Latch Mode
150%~210% of rated load current, Hiccup Mode,
(Non-Latching, Auto-Recovery)
Hiccup Mode,
(Non-Latching, Auto-Recovery)
Hiccup Mode,
(Non-Latching, Auto-Recovery)
Class I with PE* connection
3 Million Hours based on Telecordia SR-332
Input to Output
Input to Ground
Output to Ground
IEC60601-1 3rd and 3.1rd edition CB report
IEC60601-1 edition 3.1rd (2012), EN60601-1 (2006) +
A11 + A1 + A12, CAN/CSA-C22.2 NO. 60601-1:14,
ANSI/AAMI ES60601-1:2005/(R)2012
IEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-
2008, GB17625.1-2003
MDD Directive 93/42/EEC
RoHS Directive 2011/65/EU Compliant
4000 Vac
1500 Vac
500 Vac
4
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 02)