MDS-040APS24B Datasheet, PDF (3/10 Pages) Delta Electronics, Inc. – MDS Medical AC-DC Open Frame
T``E`CHNICAL DATASHEET
MDS Medical AC-DC Open Frame
24Volt, 40Watt / MDS-040APS24 B
(a) 115V
Ripple & Noise measurement circuit
Fig. 2 Ripple & Noise example, 20MHz BW
(b) 230V
PSU
5cm
0.47uF Ceramic
10uF Electrolytic
Load
Scope
Mechanical
Test Fixture
Dimensions(W x L x H)
Weight (typ.)
Terminal
Environment
Surrounding Air Temperature
Power De-rating
Operating Humidity
Operating Altitude
Shock Test (Non-Operating)
Vibration (Operating)
Input
Output
50.8 x 76.2 x 23 mm (2 x 3 x 0.91 inch)
80grams (2.88ounces)
JST 2P Connector: B2P3-VH(LF)(SN)
JST 4P Connector: B4P-VH(LF)(SN)
Operating
Storage
-10°C to +70°C
-40°C to +85°C
-10°C to +50°C 100% load
50°C to 70°C de-rate power by 2.5% / °C, See Fig. 3
10-95% RH (Non-Condensing)
3,000 meters (9842.52 feet)
50G, 11ms, 3 shocks for each direction
5-500Hz, 2.09Grms, 20 minute for each three axis
3
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 02)