English
Language : 

8205-10 Datasheet, PDF (4/4 Pages) ASM GmbH – MICRO HiCORDER

8205-10 and 8206-10 Common Specifications
Recording method 60mm (2.36in) amplitude (1 DIV = 10mm/0.39in), heat-sensitive recording
Display
Sampling rate
Level meter and scaling values by LCD bar graph, plus other
setting information
100S/s (sampling period : 10ms fixed)
Recording time axis accuracy Within ±0.5%.
Environmental conditions Operation: +5°C (41°F) to +40°C (104°F), 35% to 80% rh
(non-condensating)
Storage: -10°C (14°F) to +50°C (122°F), within 80% rh
Applicable
standards
Power
Safety: EN61010
EMC: EN61326, EN61000-3-2, EN61000-3-3
100 to 240V AC (auto selecting) at 50/60 Hz, or 9.5 to 14V DC
Maximum power consumption 30 VA maximum (AC or DC operation)
Dimensions and mass 250mm (9.4in) × 122mm (4.8in) × 93.5mm (3.66in), 1.2kg (42.33oz)
Included
Accessories
RECORDING PAPER 9235 × 1, Roll paper holder × 2, Power cord × 1,
CONNECTION CORD 9257 × 1, CARRYING CASE 9344 × 1
■ External appearance and dimensions (for both the 8205-10 and 8206-10)
250 mm (9.4 in)
Input channels
8205-10 (Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh ; accuracy guaranteed for 1 year)
One channel, AC or DC voltage, or one channel AC current.
(Simultaneous recording not available, one channel may be either type, input is voltage isolated.)
8206-10 (Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh ; accuracy guaranteed for 1 year)
One channel AC voltage and one channel AC current.
(Simultaneous recording based on two channel alternating sampling, for commercial power lines, input is voltage isolated.)
0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50, 100, 200, or 500V AC/DC. In DC mode, the zero 100, 200, or 500V AC, Max. input voltage: 600V rms. Magnified display from
Voltage
measurement ranges
* Resolution = 400points/ range
position can be set in steps of 20% of the range.
Max. input voltage: 500V rms. Accuracy: ±2% of the range. (ACV/ 45Hz to 66Hz)
Frequency characteristic: 20Hz to 30kHz/ +3 dB to -3 dB
Max. rated voltage to earth: 500V rms (the maximum voltage that can be applied be-
+25% to -35% of the range. Center position of the magnified display can be set
in 10% steps from +20% to -20%. Accuracy: ±2% of the range. (45Hz to 66Hz)
Frequency characteristic: 30 Hz to 30 kHz/ +0.5 dB to -3 dB
Max. rated voltage to earth: 600V rms (the maximum voltage that can be applied be-
tween input channel and chassis without damage)
tween input channel and chassis without damage)
Combination accuracy with the clamp-on sensor (% of full scale value)
Current
10, 20, 50, or 100A AC (with CLAMP ON SENSOR 9650).
10, 20, 50, 100, 200, or 500A AC (with CLAMP ON SENSOR 9651).
Range 10A 20A 50A 100A 200A
measurement ranges 10, 20, 50, 100, 200, 500 or 1000A AC (with CLAMP ON SENSOR 9668). 9650 ±3.8% ±3.65% ±3.56% ±3.53% -
* Resolution = 400points/ range Bar graph displays from 0% to 110% of range.
Frequency characteristic is determined by the clamp on sensor.
9651
9668
±5.0% ±4.25% ±3.8% ±3.65% ±3.58%
±8.0% ±6.5% ±5.6% ±5.3% ±5.15%
500A
-
±3.53%
±5.06%
1000A
-
-
±6.03%
Rectification system
Input resistance
True RMS.
Voltage input: approx. 1MΩ. Clamp input: 1Ω ±10%.
Ordering Information
MICRO HiCORDER 8205-10 (1-channel recording)
MICRO HiCORDER 8206-10 (2-channel simultaneous recording)
The CLAMP ON SENSOR 9650,9651 and 9668 for current measurement are not
supplied with the MICRO HiCORDERs 8205-10 and 8206-10. To measure current,
an optional CLAMP ON SENSOR must be purchased separately.
■ Option Specifications (sold separately)
Options
CONNECTION CORD
9326
(1.7m/ 5.58ft, for 8205 and
8205-10 voltage input)
Supplied as standard
CONNECTION CORD
9257
(1.3m/ 4.27ft, voltage input,
Supplied as standard)
RECORDING PAPER 9236-01
(Climate-resistant, 15m/ 49ft roll, 10
rolls/ set)
CARRYING CASE 9344 RECORDING PAPER 9235
Holds optional accessories (15m (49.22ft) roll, 10 rolls/
(Supplied as standard)
sets; one roll only, provided
with recorder)
CLAMP ON SENSOR 9650
CLAMP ON SENSOR 9651
CLAMP ON SENSOR 9668
Individual Specifications
9650
9651
9668
Rated current
Primary: 100A AC, Secondary: 100mA AC
Primary: 500A AC, Secondary: 500mA AC
Primary: 1000A AC, Secondary: 1000mA AC
Secondary current Combined accuracy with recorder depends on recorder Combined accuracy with recorder depends on recorder Combined accuracy with recorder depends on recorder
amplitude accuracy accuracy.
accuracy.
accuracy.
Frequency characteristic ±8% or better from 40Hz to 1kHz
±3% or better from 40Hz to 1kHz
±3% or better from 40Hz to 1kHz
Maximum input current 130A continuous at 45 to 66Hz
600A continuous at 45 to 66Hz
1000A continuous at 45 to 66Hz
Application circuit voltage 300V rms AC or less (insulated conductor)
600V rms AC or less (insulated conductor)
600V rms AC or less (insulated conductor)
Diameter of measurable
conductor
ø 15mm (0.59 in) or less
ø 46mm (1.81 in) or less
ø 55mm (2.17 in) or less,
80mm (3.15 in) × 20mm (0.79 in) busbar
Dimensions and mass
Approx. W 46mm (1.81 in) × H 135mm (5.31 in) × D 21mm (0.83 in) Approx. W 77mm (3.03 in) × H 151mm (5.94 in) × D 42mm (1.65 in) Approx. W 46mm (1.81 in) × H 135mm (5.31 in) × D 21mm (0.83 in)
Cable length: 3m (9.84 ft), mass: 230g (8.1 oz)
Cable length: 3m (9.84 ft), mass: 360g (12.7 oz)
Cable length: 3m (9.84 ft), mass: 230g (8.1 oz)
DISTRIBUTED BY
HEAD OFFICE :
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 / FAX +81-268-28-0568
E-mail: os-com@hioki.co.jp
HIOKI USA CORPORATION :
6 Corporate Drive, Cranbury, NJ 08512 USA
TEL +1-609-409-9109 / FAX +1-609-409-9108
E-mail: hioki@hiokiusa.com
Shanghai Representative Office :
1904, Shanghai Times Square Office
93 Huaihai Zhong Road
Shanghai, 200021, P.R.China
TEL +86-21-6391-0090, 0092
FAX +86-21-6391-0360
E-mail: info@hioki.cn
All information correct as of Mar. 5, 2007. All specifications are subject to change without notice.