English
Language : 

GCS265 Datasheet, PDF (5/8 Pages) XP Power Limited – AC-DC Power Supplies
GCS265 Series
AC-DC Power Supplies
EMC: Emissions
Phenomenon
Conducted
Radiated
Harmonic Fluctuations
Standard
EN55011/22
EN55011/22
EN61000-3-3
Test Level
Class B
Class A
Criteria
Notes & Conditions
EMC: Immunity
Phenomenon
Low Voltage PSU EMC
Harmonic Current
Radiated
EFT
Surges
Conducted
Dips and Interruptions
Standard
EN61204-3
EN61000-3-3
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN55024
(100 VAC)
EN55024
(240 VAC)
EN60601-1-2
(100 VAC)
EN60601-1-2
(240 VAC)
Test Level
High severity level
Class A
Class C
3
3
Installation class 3
3
Dip >95% (0 VAC), 8.3ms
Dip 30% (70 VAC), 416ms
Dip >95% (0 VAC), 4160ms
Dip >95% (0 VAC), 10.0ms
Dip 30% (168 VAC), 500ms
Dip >95% (0 VAC), 5000ms
Dip >95% (0 VAC), 10.0ms
Dip 60% (40 VAC), 100ms
Dip 30% (70 VAC), 500ms
Dip >95% (0 VAC), 5000ms
Dip >95% (0 VAC), 10.0ms
Dip 60% (96 VAC), 100ms
Dip 30% (168 VAC), 500ms
Dip >95% (0 VAC), 5000ms
Criteria
as below
A
A
A
A
A
B
B
A
B
B
A
A
A
B
A
A
A
B
Notes & Conditions
All models
> 70 W
Derate Output Power to 120 W
Safety Approvals
Safety Agency
CB Report
UL
TUV
CE
Equipment Protection Class
Primary to Secondary
Primary to Earth
Secondary to Earth
Safety Standard
IEC60950-1:2005 Ed 2
IEC60601-1 Ed 3 Including Risk Management
UL60950-1 (2007), CSA 22.2 No.60950-1-1:08
ANSI/AAMI ES60601-1:2005 & CSA C22.2, No.60601-1:08
EN60950-1:2006
EN60601-1/A12:2006
LVD & RoHS
Class I & Class II
Notes & Conditions
Information Technology
Medical
Information Technology
Medical
Information Technology
Medical
See safety agency conditions of acceptibility for details
Means of Protection
2 x MOPP (Means of Patient Protection)
1 x MOPP (Means of Patient Protection)
1 x MOPP (Means of Patient Protection)
Category
IEC60601-1 Ed 3
55
www.xppower.com