English
Language : 

EML30 Datasheet, PDF (3/4 Pages) XP Power Limited – AC-DC Power Supplies
EML30 Series
AC-DC Power Supplies
EMC: Emissions
Phenomenon
Conducted
Radiated
Harmonic Current
Voltage Flicker
EMC: Immunity
Phenomenon
ESD
Radiated
EFT
Surges
Conducted
Magnetic Fields
Dips and Interruptions
Safety Approvals
Safety Agency
CB Report
UL
TUV
Primary to Secondary
Standard
EN55022/11:2010
EN55022/11:2010
EN61000-3-2
EN61000-3-3
Test Level
Class B
Class B
Criteria
Notes & Conditions
Class A
Standard
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN60601-1-2 (100 VAC)
EN60601-1-2 (240 VAC)
Test Level
3
10 V/m
3
3
10 Vrms
30 A/m
70% UT (70 VAC) for 500 ms
40% UT (40 VAC) for 100 ms
<5% UT (0 VAC) for 10 ms
<5% UT (0 VAC) for 5000 ms
70% UT (168 VAC) for 500
40% UT (96 VAC) for 100 ms
<5% UT (0 VAC) for 10 ms
<5% UT (0 VAC) for 5000 ms
Criteria
A
A
A
A
A
A
A
A
A
B
A
A
A
B
Notes & Conditions
6 kV contact, 8 kV air discharge
80% mod
Derate output power to 10 W
Safety Standard
IEC60601-1 Ed 3 Including Risk Management
ANSI/AAMI ES60601-1:2005 & CSA C22.2, No.60601-1:08
EN60601-1/A11:2011
Notes & Conditions
Medical
Medical
Medical
Means of Protection
2 x MOPP (Means of Patient Protection)
Category
IEC60601-1 Ed 3
3
www.xppower.com