English
Language : 

EML30 Datasheet, PDF (1/4 Pages) XP Power Limited – AC-DC Power Supplies
EML30 Series
AC-DC Power Supplies
30 Watts
• Compact Size
• Medical Approvals
• Single Outputs from 3.3 to 48 V
• PCB Mount, Open Frame & Chassis Mount
• Encapsulated PCB & Chassis Mount
• Class II
• Peak Load Capability
The EML30 is a series of open frame and encapsulated AC-DC single output power supplies
designed for medical applications. The series provides a number of flexible mechanical options
including PCB mount, open frame, chassis mount with screw terminals and a DIN Rail option. With
approvals to world-wide medical safety standards, compliance with class B for both conducted and
radiated emissions and a 130%, 30s peak load capability, these class II isolation parts benefit
system designers with easy integration into the latest healthcare products and applications.
Dimensions:
EML30:
( P ): 2.96 x 1.36 x 1.05” (75.2 x 34.6 x 26.7 mm)
( T ): 3.46 x 1.36 x 1.00” (87.9 x 34.6 x 25.4 mm)
( E ): 3.10 x 1.50 x 1.10” (78.7 x 38.1 x 27.9 mm)
( S): 3.78 x 1.57 x 1.12” (96.0 x 40.0 x 28.5 mm)
Models & Ratings
Output Power
20 W
30 W
30 W
30 W
30 W
30 W
30 W
30 W
Output Voltage
3.3 VDC
5.0 VDC
9.0 VDC
12.0 VDC
15.0 VDC
24.0 VDC
36.0 VDC
48.0 VDC
Nominal
6.00 A
6.00 A
3.33 A
2.50 A
2.00 A
1.25 A
0.83 A
0.62 A
Output Current
Peak (1)
7.80 A
7.80 A
4.33 A
3.25 A
2.60 A
1.63 A
1.08 A
0.81 A
Model Number (2,3)
EML30US03
EML30US05
EML30US09
EML30US12
EML30US15
EML30US24
EML30US36
EML30US48
Notes
1. Peak load lasting <30 s with a maximum duty cycle of 10%, average output power not to exceed nominal.
2. Add suffix to model number to define type: add ‘-P’ for PCB mount, add ‘-T’ for chassis mount, add ‘-E’ for encapsulated, add ‘-S’ for screw terminals.
3. A Screw terminal version (-S) is available with DIN clip attached, add suffix ‘D’, e.g. EML30US24-SD, DIN rail mounting kit is available as a seperate item,
order code ECL25/30 DIN CLIP.
Summary
Characteristic
Input Voltage Range
No Load Input Power
Efficiency
Operating Temperature
EMC
Safety Approvals
Minimum
Typical
Maximum
Units
Notes & Conditions
85
264
VAC
Derate from 100% at 90 VAC to 90% at 85 VAC
120
370
VDC
0.3
W
79
88
%
-20
+70
°C
Derate linearly from 100% at +50 °C to 50% at +70 °C
EN55022 Level B Conducted & Radiated, EN61000-4, EN61000-3, EN60601-1-2
EN60601-1, ANSI/AAMI ES60601-1. IEC60601-1
1
www.xppower.com