HPU1K5 Datasheet, PDF (1/4 Pages) XP Power Limited – 1500 Watts HPU1K5 Series | |||
|
1500 Watts
HPU1K5 Series
xppower.com
⢠Medical Safety Approvals ( M Versions)
⢠Variable Fan Speed To Reduce Audible Noise
⢠20 °C to +70 °C Operation
⢠AC OK, DC OK, Inhibit & 5 V Standby Supply
⢠Fault & Overtemperature Signals
⢠SEMI F47 Compliant
⢠3 Year Warranty
Specification
Input
Input Voltage
Input Frequency
Input Current
Inrush Current
Power Factor
Earth Leakage Current
Input Protection
⢠85-264 VAC, see derating curve
⢠47-63 Hz
⢠13 A/6.5 A typical at 115/230 VAC
⢠35 A maximum at 264 VAC
⢠>0.9
⢠1.1 mA max at 264 VAC 60Hz, <500 µA at
264 VAC max, <300 µA at 130 VAC (-M)
⢠Internal T20 A/250 VAC fuse in line and
neutral
Output
Output Voltage
⢠See model table
Output Voltage Trim
⢠Via potentiometer or external voltage,
see model tables
Initial Set Accuracy ⢠±1% of nominal with 50% load
Minimum Load
⢠No minimum load required
Line Regulation
⢠±0.5% maximum
Load Regulation
⢠V1: ±0.5%, V2: ±5%
Start Up Delay
⢠1 s typical
Over/Undershoot
⢠0.5% typical
Transient Response
⢠4% deviation, recovery to within 2% in
500 µs for 50-75-50% load change
Ripple & Noise
⢠24-48 V models: 1% max pk-pk
12 V models: 2% max pk-pk
V Standby: 3% max pk-pk,
20 MHz bandwidth
Overvoltage Protection ⢠115-140% of V1 nominal, recycle input
AC to reset
Overtemperature
Protection
⢠Protects the unit against overtemperature.
Auto restart
Overcurrent Protection ⢠110 - 140% V1, V Standby power limited
Short Circuit Protection ⢠Continuous, trip and restart (hiccup mode)
Temperature
Coefficient
⢠0.02%/°C (after 20 minute warm up)
Remote Sense
⢠Compensates for 0.5 V total drop
Current Share
⢠Share upto 8 units maximum, units
share current within 10% of each other
at full load.
General
Efficiency
Isolation
Switching Frequency
Power Density
Signals
MTBF
⢠90% typical
⢠4000 VAC Input to Output 2 x MOPP,
1500 VAC Input to Ground 1 x MOPP
500 VDC Output to Ground
⢠70 kHz (PFC), 130 kHz (main converter)
typical
⢠18 W/in3
⢠AC OK, DC OK, Inhibit, Fault
(see Signals page)
⢠470 kHrs to Telecordia SR-332 at 25 °C, GB
Environmental
Operating Temperature ⢠-20 °C to +70 °C, derate linearly from
+50 °C at 2.5 %/°C to 50% load at +70°C
Cooling
⢠Internal load dependant variable speed fans
Operating Humidity ⢠95% RH, non-condensing
Storage Temperature ⢠-40 °C to +85 °C
Operating Altitude
⢠3000 m
Shock
⢠±3 shocks in each axis (total 18 shocks)
30 g 11 ms (half sine). Compliant with
EN60068-2-27.
Vibration
⢠2 g 10-500 Hz 10 sweeps. Compliant with
EN60068-2-6.
EMC & Safety
Emissions
Immunity
Harmonic Currents
Voltage Flicker
ESD Immunity
Radiated Immunity
EFT/Burst
Surge
Conducted Immunity
Dips & Interruptions
Safety Approvals
⢠EN55011 level A conducted & radiated,
EN55022 level A conducted & radiated
⢠Compliant with EN61204-3:2000
high severity levels
⢠EN61000-3-2 class A,
EN61000-3-2 class C for loads â¥10%
⢠EN61000-3-3
⢠EN61000-4-2, level 3, Perf Criteria A
⢠EN61000-4-3, level 3, Perf Criteria A
⢠EN61000-4-4, level 3, Perf Criteria A
⢠EN61000-4-5, installation class 3
Perf Criteria A, SEMI F47
⢠EN61000-4-6, level 3, Perf Criteria A
⢠EN61000-4-11, 30% 10 ms, 60% 100 ms,
100% 5000 ms, Perf Criteria A, B, B,
EN60601-1-2, 30% 500 ms, 60% 100 ms,
100% 10 ms, 100% 5000 ms, Perf Criteria
A, A, A, B
⢠EN60601-1, ANSI/AAMI ES60601-1,
CSA22.2 No.60601-1 per cUL, Including
Risk Management M Versions, EN60950,
UL60950
|
▷ |