English
Language : 

AFE4490_14 Datasheet, PDF (67/96 Pages) Texas Instruments – Integrated Analog Front-End for Pulse Oximeters
www.ti.com
AFE4490
SBAS602G – DECEMBER 2012 – REVISED JUNE 2014
Figure 108. ADCRSTENDCT2: ADC Reset 2 End Count Register (Address = 1Ah, Reset Value = 0000h)
23
22
21
20
19
18
17
16
15
14
13
12
0
0
0
0
0
0
0
0
ADCRSTENDCT2[15:0]
11
10
9
8
7
6
5
4
3
2
1
0
ADCRSTENDCT2[15:0]
This register sets the end position of the ADC2 reset conversion signal.
Bits D[23:16]
Bits D[15:0]
Must be '0'
ADCRSTENDCT2[15:0]: ADC RESET 2 end count
The contents of this register can be used to position the end of the ADC reset conversion.
Refer to the Using the Timer Module section for details.
Figure 109. ADCRSTSTCT3: ADC Reset 3 Start Count Register (Address = 1Bh, Reset Value = 0000h)
23
22
21
20
19
18
17
16
15
14
13
12
0
0
0
0
0
0
0
0
ADCRSTSTCT3[15:0]
11
10
9
8
7
6
5
4
3
2
1
0
ADCRSTSTCT3[15:0]
This register sets the start position of the ADC3 reset conversion signal.
Bits D[23:16]
Bits D[15:0]
Must be '0'
ADCRSTSTCT3[15:0]: ADC RESET 3 start count
The contents of this register can be used to position the start of the ADC reset conversion.
Refer to the Using the Timer Module section for details.
Figure 110. ADCRSTENDCT3: ADC Reset 3 End Count Register (Address = 1Ch, Reset Value = 0000h)
23
22
21
20
19
18
17
16
15
14
13
12
0
0
0
0
0
0
0
0
ADCRSTENDCT3[15:0]
11
10
9
8
7
6
5
4
3
2
1
0
ADCRSTENDCT3[15:0]
This register sets the end position of the ADC3 reset conversion signal.
Bits D[23:16]
Bits D[15:0]
Must be '0'
ADCRSTENDCT3[15:0]: ADC RESET 3 end count
The contents of this register can be used to position the end of the ADC reset conversion
signal (default value after reset is 0000h). Refer to the Using the Timer Module section for
details.
Copyright © 2012–2014, Texas Instruments Incorporated
Product Folder Links: AFE4490
Submit Documentation Feedback
67