US-TUVR-4428 Datasheet, PDF (8/10 Pages) SL Power Electronics – IEC SYSTEM FOR MUTUAL RECOGNITION OF TEST CERTIFICATES FOR ELECTRICAL...
Issue Date:
Correction 1
2008-05-16
2009-04-10
Page 4 of 5
Report Reference #
CF2.2
CF2.3
CF2.4
CF2.5
CF2.6
CF2.7
CF2.8
CF2.9
CF3
CF3.1
The output circuits have not been evaluated for --
direct patient connection (Type B, BF or CF).
The Temperature Test was performed in an --
ambient of 25 °C and were calculated for a
raised ambient of 40°C. In some instances, the
temperature test was repeated in an ambient of
40°C to validate the calculation.
The power supply was evaluated as Reinforced --
insulation between primary and secondary;
basic insulation between primary to heatsink;
and operational insulation between secondary to
heatsink.
This power supply has been evaluated as Class I --
equipment, continuous operation, and has not
been evaluated for use in the presence of a
flammable anesthetic mixture with air, oxygen,
or nitrous oxide.
The output has been evaluated and tested for --
operator accessible part. Under normal and
single fault conditions, the outputs do not exceed
25 V ac or 60 V dc.
The product was evaluated for desktop and/or --
portable use only. Additional evaluation may be
required as part of end product investigation to
insure compliance of the product as mobile
device.
Leakage current measurements must be
--
performed on the combination power supply and
the end-use product.
Installation instructions and end product
--
markings are the responsibility of the end-use
product manufacturer.
At 90V~ input, the 12V model output is derated --
to 12Vdc, 8.33A to comply with the temperature
limit. Above 90V~ input, the 12V model complies
with the temperature limit with a rated output
12Vdc, 9.165A
Earth leakage current was measured using Figs --
10 and 11 of UL60601-1.
E145177-A9-UL-1
TRF No.: IEC60601_1C
Underwriters Laboratories Inc.