US-TUVR-4428 Datasheet, PDF (7/10 Pages) SL Power Electronics – IEC SYSTEM FOR MUTUAL RECOGNITION OF TEST CERTIFICATES FOR ELECTRICAL...
Issue Date:
Correction 1
2008-05-16
2009-04-10
Page 3 of 5
Report Reference #
E145177-A9-UL-1
CD1.0
CD1.1
Additional Information
The schematics are kept on file at the CB Testing Laboratory mentioned in the first page of this test
report, and can be provided by the applicant upon request by CBTLs/NCBs.
All tests were conducted at the SMT facility.
CE1.0
CE1.1
CE1.2
CE1.3
CE1.4
CE1.5
CE1.6
CE1.7
CE1.8
Technical Considerations
The product was investigated to the following
additional standards:
UL 60601-1, 1st Edition, 2006-04-26 (includes
National Differences for USA), CAN/CSA-C22.2
No. 601.1-M90 (R2005) (includes National
Differences for Canada), UL 60601-1, 1st
Edition, 2006-04-26 (includes National
Differences for USA)
The product was not investigated to the following Clause 36, Electromagnetic Compatibility (IEC
standards or clauses:
601-1-2), Clause 48, Biocompatibility (ISO
10993-1), Clause 52.1, Programmable
Electronic Systems (IEC 601-1-4),
The product is Classified only to the following
hazards:
Casualty, Fire , Shock
The degree of protection against harmful ingress Ordinary
of water is:
The following accessories were investigated for ,
use with the product:
The mode of operation is:
Continuous
Software is relied upon for meeting safety
No
requirements related to mechanical, fire and
shock:
The product is suitable for use in the presence of No
a flammable anesthetics mixture with air or
oxygen or with nitrous oxide:
CF1.0
CF1.1
CF2.0
CF2.1
Engineering Conditions of Acceptability
For use only in or with complete equipment where the acceptability of the combination is
determined by Underwriters Laboratories Inc.
When installed in an end-product, consideration must be given to the following:
This component has been judged on the basis of --
the required spacings in the First Edition of the
Standards for Medical Electrical Equipment, Part
1: General Requirements for Safety, UL
60601-1, which covers the end use product for
which the component is designed.
The component is provided with a plastic
--
enclosure and an appliance inlet
TRF No.: IEC60601_1C
Underwriters Laboratories Inc.