English
Language : 

FFUS15-1G1IO Datasheet, PDF (2/5 Pages) SICK AG – FLOW SENSORS
FFUS15-1G1IO | FFU
FLOW SENSORS
A
B
C
D
E
F
H
I
J
K
L
M
N
O
P
Q
R
S
T
Ordering information
Type
FFUS15-1G1IO
Other models and accessories www.sick.com/FFU
Illustration may differ
Part no.
6041249
Features
Medium
Nominal width measuring tube
Process temperature
Process pressure
Fluids
DN 15
0 °C ... +80 °C
Max. 16 bar
Performance
Minimim flow
Maximum flow
Inlet zone
Output zone
Conductivity
Accuracy of sensor element
Reproducibility
Resolution
≤ 0.9 l/min
≤ 36 l/min
30 cm
5 cm
No limitation
2 % (of final value) 1)
≤ 0.5 %
0.006 l/min
1) Reference condition: Water, zero gas, measuring tube completely full, no cavitation, temperature of medium 20 °C, ambient temperature 20 °C ... 25 °C, compli-
ance with inlet and outlet zones, warm-up time for electronics: 30 min.
Electronics
Supply voltage
Residual ripple
Power consumption
Initialization time
Protection class
Electrical connection
18 V DC ... 30 V DC 1)
≤ 5 Vpp 2)
≤ 180 mA 3)
≤5s
III
Round connector M12 x 1, 5-pin
1) All connections are polarity protected. All outputs are overload and short-circuit protected.
2) May not exceed or fall below Uv tolerances.
3) Without load.
4) There are 100 mA for each output pnp and npn available.
5) Analog output and display.
2
FLUID SENSORS | SICK
Online data sheet | 2017-02-08 13:58:08
Subject to change without notice