English
Language : 

DSA8300 Datasheet, PDF (1/26 Pages) List of Unclassifed Manufacturers – Digital Serial Analyzer Sampling Oscilloscope
DSA8300 Datasheet
Digital Serial Analyzer Sampling Oscilloscope
The DSA8300 is a state-of-the-art Equivalent Time Sampling Oscilloscope
that provides the highest fidelity measurement and analysis capabilities for
Communications Signal Analysis, Serial Data Network Analysis, and Serial
Data Link Analysis applications.
Key performance specifications
Very low time base jitter:
425 fs typical on up to 8 simultaneously acquired channels
<100 fs on up to 6 channels with 82A04B phase reference module
Industry’s highest vertical resolution – 16 bit A/D
Electrical resolution: <20 µV LSB (for 1 v full range)
Optical resolution from <20 nW for the 80C07B (1 mW full range) to
<0.6 µW for the 80C10C (30 mW full range)
Optical bandwidths to >80 GHz
Electrical bandwidths to >70 GHz
Over 120 automated measurements for NRZ, RZ, and pulse signal
types
Automated mask testing with over 80 industry-standard masks
Key features
With today’s sampling module portfolio, the DSA8300 supports up to
8 simultaneously acquired signals
A wide variety of optical, electrical, and accessory modules to support
your specific testing requirements
Optical modules
Fully integrated optical modules that support all standard optical
data rates from 155 Mb/s to 100 Gb/s
Certified optical reference receivers support specified requirements
for standards-mandated compliance testing
Optical bandwidths to >80 GHz
High optical sensitivity, low noise, and wide dynamic range of the
optical sampling modules allows accurate testing and
characterization of short-reach to long-haul optical communications
standards
Fully calibrated clock recovery solutions – no need to manually
calibrate for data pick-off losses
Calibrated extinction ratio measurements ensure repeatability of
extinction ratio measurements to <0.5 dB among systems with
modules with this factory calibration option
Electrical modules
Very low-noise electrical samplers (280 μV at 20 GHz, 450 μV at
60 GHz, typical)
Selectable bandwidths 1 allow the user to trade-off sampler
bandwidth and noise for optimal data acquisition performance
Remote samplers 2 or compact sampling extender module cables
support minimal signal degradation by allowing the sampler to be
located in close proximity to the device under test
World’s highest-performance integrated TDR (10 ps typical step
rise time) supports exceptional impedance discontinuity
characterization and high dynamic range for s-parameter
measurements to 50 GHz
1 With 80E07B, 80E08B, 80E09B, 80E10B, 80E11, and 80E11X1 modules.
2 80E07B, 80E08B, 80E09B and 80E10B only.
www.tektronix.com 1