English
Language : 

02S30R-MSOTS3 Datasheet, PDF (2/3 Pages) Rosenberger Hochfrequenztechnik GmbH & Co. KG – Calibration Kit
Technical Data Sheet
RPC-2.92
Calibration Kit
Plug
02S30R-MSOTS3
Electrical data
Frequency range
DC to 40.0 GHz
Thru
Return loss
≥ 32 dB, DC to 4 GHz
≥ 30 dB, 4 GHz to 26.5 GHz
≥ 28 dB, 26.5 GHz to 40 GHz
Open
Error from nominal phase1
≤ 1.5°, DC to 4 GHz
≤ 4.0°, 4 GHz to 26.5 GHz
≤ 5.0°, 26.5 GHz to 40.0 GHz
Short
Error from nominal phase2
≤ 1.5°, DC to 4 GHz
≤ 4.0°, 4 GHz to 26.5 GHz
≤ 5.0°, 26.5 GHz to 40.0 GHz
Load
Return loss
DC Resistance
Power handling
≥ 40.0 dB, DC to 4 GHz
≥ 28.0 dB, 4 GHz to 26.5 GHz
≥ 25.0 dB, 26.5 GHz to 40.0 GHz
50 Ω ± 0.5 Ω
≤ 0.5 W
1 The nominal phase is defined by the Offset Delay, the Offset Loss and the Fringing Capacitances
2 The nominal phase is defined by the Offset Delay, the Offset Loss and the Short Inductance
Mechanical data
Mating cycles
Maximum torque
Recommended torque
Gauge
≥ 500
1.70 Nm
0.90 Nm
0.00 mm to 0.08 mm
General standard definitions
For proper operation the vector network analyzer (VNA) needs a model describing the electrical behaviour
of this calibration standard. The different models, units, and terms used will depend on the VNA type and they
will have to be entered into the VNA. All values are based on typical geometry and plating.
Thru
Offset Zo / Impedance / Zo
Offset Delay
Length (electrical) / Offset Length
Offset Loss
Loss
50 Ω
83.057 ps
24.90 mm
2.70 GΩ/s
0.0195 dB/ GHz
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.de
Tel. : +49 8684 18-0
Email : info@rosenberger.de
Page