English
Language : 

PXIE-5663 Datasheet, PDF (6/8 Pages) National Instruments Corporation – Vector Signal Analyzer
Vector Signal Analyzer
0 dBm Reference Level
Frequency Range
10 MHz to <30 MHz
30 MHz to <330 MHz
330 MHz to <3.0 GHz
3.0 GHz to 6.6 GHz
Note: Two - 4 dBm input tones = 200 kHz apart.
Input IP3
≥21 dBm
≥18 dBm
≥21 dBm
≥21 dBm
Dynamic Range1
Dynamic Range (Noise and Third-Order Intermodulation
Distortion (IMD3))
(Nominal)
10 –
–10 –
–30 –
Band 1 = 30 MHz to 330 MHz
Band 2 = 330 MHz to 3.0 GHz
Band 3 = 3.0 GHz to 6.6 GHz
–50 –
–70 –
–90 – Noise
–110 –
Digitizer IMD3
NI
5663
Band
1 IMD3
NI 5663
Band
3
IMD3
–130 –
NI 5663 Band 2 IMD3
–150 –
–40
–30
–20
–10
0
Input Power (dBm, per tone)
Figure 4. NI 5663 Vector Signal Analyzer Nominal Dynamic Range, 0 dBm Reference Level
10 –
–10 –
Band 1 = 30 MHz to 330 MHz
Band 2 = 330 MHz to 3.0 GHz
Band 3 = 3.0 GHz to 6.6 GHz
–30 –
–50 –
–70 –
Noise
Digitizer IMD3
–90 –
NI 5663 Band 2 IMD3
–110 –
–130
– NI 5663 Band 3 IMD3
NI 5663 Band 1 IMD3
–150 –
–60
–50
–40
–30
–20
Input Power (dBm, per tone)
Figure 5. NI 5663 Vector Signal Analyzer Nominal Dynamic Range, -20 dBm Reference Level
1Reference level allows 10 dB headroom for single-tone input signals before digitizer clipping occurs.
The dynamic range plots in the two preceding figures show nominal
performance with NI-RFSA automatic coupled settings that are optimized
for noise performance. If you use the RF attenuation manual settings, IMD3
performance can improve with minimal degradation in noise floor, thus
increasing the effective SFDR in the power per tone signal range of 10 to 0 dB
below reference level.
IF Flatness
(Typical)
IF Amplitude Flatness, 23 °C ± 5 °C
Tuned Frequency
Bandwidth
Amplitude Flatness
10 MHz to <75 MHz
5 MHz
10 MHz
±0.25 dB
±0.3 dB
75 MHz to <120 MHz
5 MHz
10 MHz
±0.4 dB
±0.6 dB
5 MHz
±0.45 dB
120 MHz to <140 MHz
10 MHz
±0.65 dB
20 MHz
±0.9 dB
5 MHz
±0.2 dB
140 MHz to <330 MHz
10 MHz
±0.4 dB
20 MHz
±0.5 dB
10 MHz
±0.2 dB
330 MHz to <6.6 GHz
20 MHz
±0.35 dB
50 MHz
±0.60 dB
Notes: RF attenuation ≥8 dB, 18 to 28 °C, with calibration correction; bandwidth centered about
tuned frequency. Typical represents the worst ripple expected for any reference level setting across
the specified frequency range.
IF Amplitude Flatness, 0 to 55 °C
Tuned Frequency
Bandwidth
Amplitude Flatness
10 MHz to <75 MHz
5 MHz
10 MHz
±0.3 dB
±0.45 dB
75 MHz to <120 MHz
5 MHz
10 MHz
±0.35 dB
±0.6 dB
5 MHz
±0.55 dB
120 MHz to <140 MHz
10 MHz
±0.85 dB
20 MHz
±1.1 dB
5 MHz
±0.35 dB
140 MHz to <330 MHz
10 MHz
±0.8 dB
20 MHz
±0.8 dB
10 MHz
±0.25 dB
330 MHz to <6.6 GHz
20 MHz
±0.4 dB
50 MHz
±0.7 dB
Notes: RF attenuation ≥8 dB, 0 to 55 °C, with calibration correction; bandwidth about tuned frequency.
Typical represents the worst ripple expected for any reference level setting across the specified
frequency range.
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)