English
Language : 

PXI-4204 Datasheet, PDF (2/4 Pages) National Instruments Corporation – DAQ for High-Voltage Measurements 16-Bit, 200 kS/s, 8 Analog Inputs
DAQ for High-Voltage Measurements
16-Bit, 200 kS/s, 8 Analog Inputs
Specifications – PXI-4204
Complete Accuracy Tables
Input Characteristics
SSH Disabled
Nominal Range (V)
Positive FS Negative FS
100
-100
50
-50
5
-5
0.5
-0.5
% of Reading
24 Hrs
1 Year
0.063
0.070
0.031
0.038
0.061
0.068
0.061
0.068
SSH Enabled
Nominal Range (V)
Positive FS Negative FS
100
-100
50
-50
5
-5
0.5
-0.5
% of Reading
24 Hrs
1 Year
0.085
0.090
0.035
0.042
0.063
0.070
0.063
0.070
Absolute Accuracy
Offset
Noise + Quantization (mV)
(mV)
Single Pt.
Averaged
± 16.8
± 9.34
± 0.92
± 10.4
± 4.68
± 0.57
± 4.7
± 0.69
± 0.40
± 4.1
± 0.49
± 0.40
Absolute Accuracy
Offset
Noise + Quantization (mV)
(mV)
Single Pt.
Averaged
± 19.0
± 9.35
± 1.02
± 12.6
± 4.70
± 0.73
± 6.8
± 0.82
± 0.60
± 6.3
± 0.66
± 0.60
Temp
Drift (%/°C)
0.0027
0.0025
0.0027
0.0027
Temp
Drift (%/°C)
0.0027
0.0025
0.0027
0.0027
Absolute Accuracy
at Full Scale
(mV)
88
30
8.5
4.8
Percent Error
at Full Scale
(%)
0.088
0.060
0.17
0.97
Absolute Accuracy
at Full Scale
(mV)
110
34
11
7.2
Percent Error
at Full Scale
(%)
0.110
0.069
0.22
1.4
Note: Accuracies are valid for measurements following an internal calibration.
Averaged numbers assume dithering and averaging of 100 single-channel
readings. Measurement accuracies are listed for operational temperatures within
±1°C of internal calibration temperature and ±10°C of external or factory
calibration temperature.
Number of Channels
PXI-4204
8 differential
Input signal ranges..................................... see accuracy tables
Resolution .................................................. 16 bits, 1 in 65,536
Sampling Rate
Maximum Sampling Rate .......................... 200 kS/s aggregate multi-channel
Module
PXI-4204
Single Channel
333 kS/s
Max per channel (scanning all channels)
Scan Rate w/SSH*
Scan Rate w/o SSH*
20 kS/s/ch
25 kS/s/ch
*SSH = Simultaneous Sample and Hold
Connector
Screw Terminals
SMB Connector
Powered On
± 100 VDC
± 15 V
Powered Off
± 100 VDC
± 15 V
Overvoltage protection
Inputs with overvoltage protection ............ CH<0..7>
Input coupling ............................................ DC
Data transfers ............................................ DMA, interrupts, programmed I/O
Transfer Characteristics
Nonlinearity
Range
± 100 V
Other Ranges
± 100 V, SSH Enabled
Other Ranges, SSH Enabled
Percent of Full Scale (Typical)
0.02%
0.01%
0.06%
0.02%
DNL
Module
PXI-4204
Typical
± 0.5 LSB
Maximum
± 1 LSB
Gain Error ................................................... See Accuracy Table
Offset Error ................................................ See Accuracy Table**
Amplifier Characteristics
Input Impedance
Module
PXI-4204
Input Configuration
Differential Input
Single-Ended
Normal Powered On
2 MΩ
1 MΩ
Powered Off
2MΩ
1MΩ
NMR (60 Hz)
PXI-4204
40 dB @ 6 Hz filter setting
CMRR
Module
PXI-4204
PXI-4204
Input Frequency Ranges
DC to 60 Hz
100 Hz to 10 kHz
CMRR
>60 dB
>40 dB
Filter Characteristics
Filter Type .................................................. 6 Hz or 10 kHz, 2-pole, Butterworth
Stability
Module Input Range Gain Temperature Coefficient Offset Temperature Coefficient
PXI-4204 ± 100 V
± 25 ppm/Co
± 385 uV/Co
Triggers
Digital Triggers
Number of triggers ............................... 2
Purpose ................................................ Start & stop trigger, gate, clock
Source .................................................. PFI0/TRIG1 (front SMB connector),
PXI_TRIG_0…PXI_TRIG_6
(PXI trigger bus)
Slope .......................................................... Positive or negative; software
selectable
Compatibility......................................... 5V/TTL
Response.............................................. Rising or falling edge
Pulse width........................................... 10 ns minimum
Impedance............................................ 10 kOhm
Coupling................................................ DC
PXI Trigger Bus
Trigger lines .......................................... 6
Star trigger............................................ 1
Bus Interface
PXI……….……………………….Master, slave
Power Requirements
Module
PXI-4204
5 VDC (±5%)
1A
Calibration
Recommended warm-up time ............. 15 minutes
External calibration interval................... 1 year
Onboard calibration reference
Level................................................ 5.000V +/- 1mV
Temperature coefficient .................. + 0.6 ppm/oC
Long-term stability .......................... 6 ppm/Sqrt(1000h)
Physical
Dimensions (not including
connectors).......................................... 16.0 by 10.0 cm by 2.0 cm
(6.3 x 3.9 x 0.79 in)
Analog input signal connector.................. 16 x 1 minicombicon, 3.81 mm pitch
Analog input signal mating connector ..... 16 x 1 minicombicon, 3.81 mm pitch,
28-16 AWG signal wire
**Gain Error = Actual Input Voltage * % of Reading
2 National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • info@ni.com • ni.com