English
Language : 

PCI-5142 Datasheet, PDF (14/77 Pages) National Instruments Corporation – 14-Bit 100 MS/s Digitizer with Onboard Signal Processing
Specification
Sample Clock
Jitter, Typical
Value
≤1 ps rms (100 Hz to 100 kHz)
≤2 ps rms (100 Hz to 1 MHz)
Timebase
Frequency
100 MHz
Timebase
Accuracy
Not Phase-Locked
to Reference Clock
Phase-Locked to
Reference Clock
±25 ppm
Equal to the Reference
Clock accuracy
Sample Clock ±1 Sample Clock period
Delay Range
Sample Clock
Delay
Resolution
10 ps
External Sample Clock
Sources
NI PXI-5142
NI PCI-5142
CLK IN (front panel CLK IN (front panel
SMB connector)
SMB connector)
PXI Star Trigger
(backplane
connector)
Comments
Includes the effects of
the converter aperture
uncertainty and the clock
circuitry jitter. Excludes
trigger jitter.
—
—
—
—
—
NI PXI/PCI-5142 Specifications
14
ni.com