PCI-5142 Datasheet, PDF (1/77 Pages) National Instruments Corporation – 14-Bit 100 MS/s Digitizer with Onboard Signal Processing
NI PXI/PCI-5142 Specifications
14-Bit 100 MS/s Digitizer with Onboard Signal Processing
̭͈Ρ΅νιϋΠ͉ͅȂ඾ུࢊβȜΐ̳̞̀ͦ͘͘͜܄ȃ
This document lists the specifications for the NI PXI/PCI-5142 (NI 5142)
high-speed digitizer. Unless otherwise noted, the following conditions were
used for each specification:
• All filter settings
• All impedance selections
• Sample clock set to 100 MS/s
Typical values are used to define an average unit measured at ambient
temperatures of 15 °C to 35 °C. Specifications are subject to change
without notice. For the most recent NI 5142 specifications, visit ni.com/
manuals.
To access NI 5142 documentation, navigate to Start»All Programs»
National Instruments»NI-SCOPE»Documentation.
Hot Surface If the NI 5142 has been in use, the device or the shield may exceed safe
handling temperatures and may cause burns. Allow the NI 5142 to cool before touching the
shield or removing the device from the chassis or PC. Refer to the Environment section for
operating temperatures.
Contents
Vertical.................................................................................................... 2
Analog Input (Channel 0 and Channel 1) ........................................ 2
Horizontal................................................................................................ 13
Sample Clock ................................................................................... 13
Phase-Locked Loop (PLL) Reference Clock................................... 16
CLK IN (Sample Clock and Reference Clock Input,
Front Panel Connector) ................................................................. 16
CLK OUT (Sample Clock and Reference Clock Output,
Front Panel Connector) ................................................................. 17