English
Language : 

HPU100 Datasheet, PDF (2/2 Pages) HITRON ELECTRONICS CORPORTION – UNIVERSAL AC INPUT HARMONIC CORRECTION AC-DC 19" COMPATIBLE EUROCARD MULTI-OUTPUT 100 WATTS SWITCHING MODE POWER SUPPLIES
HPU100 SERIES
100W Desk top type medical power supplies for Home health care applications.
Safety Specifications:
Sym.
Parameter
Vps Dielectric Withstanding Voltage
for Primary to secondary
Vpg
Dielectric Withstanding Voltage
for Primary to Ground
Ris Isolation Resistance
CISPR EMI requirements for CISPR-11
FCC EMI requirements for FCC PART-18
Test Conditions
Primary to secondary
Primary to ground
Test Voltage = 2100VDC
Vin=230VAC
Vin=120VAC
Min. Typ. Max. Unit
5600
VDC
2800
50
B
B
VDC
M
CLASS
CLASS
Output Voltage And Current Rating Chart (Single Output) :
Model Number
HPU100-105
HPU100-106
HPU100-107
HPU100-108
HPU100-109
HPU100-111
Output Voltage
11 ~ 13 VDC
13 ~ 16 VDC
16 ~ 21 VDC
21 ~ 27 VDC
27 ~ 33 VDC
40 ~ 48 VDC
Output Current
9.09 ~ 7.69 A
7.69 ~ 6.25 A
6.25 ~ 4.76 A
4.76 ~ 3.70 A
3.70 ~ 3.03 A
2.50 ~ 2.08 A
Total Regulation ①
5%
4%
4%
4%
3%
3%
The output voltage under 30V had been approved by TUV/PSE.
① The total regulation on each model is required to use AWG#18×3C+ AWG#16×3C/ 4FT output cable.
The regulation will be changed by modified output cable.
Maximum
Output Power
100W
100W
100W
100W
100W
100W
Mechanical Specifications:
Note:
1. Dimensions are shown in inches or mm.
2. Weight: 460-800gs approx.
3. Optional output connector: