English
Language : 

HBU60 Datasheet, PDF (2/2 Pages) MPS Industries, Inc. – 60W Open frame type medical power supplies for health care applications
HBU60 SERIES
60W Open frame type medical power supplies for health care applications
Safety specifications:
Sym.
Parameter
Vps
Dielectric Withstanding Voltage
for Primary to secondary
Vpg
Dielectric Withstanding Voltage
for Primary to Ground
Ris Isolation Resistance
CISPR EMI requirements for CISPR-11
FCC EMI requirements for FCC PART-18
T e s t Conditions
Primary to secondary
Primary to ground
Test Voltage=500VDC
Vin=220VAC
Vin=110VAC
Min. Typ. Max. Unit
5656
VDC
2828
50
B
B
VDC
M
CLASS
CLASS
Output voltage and current rating chart (Single output) :
Model Number
HBU60- 105
HBU60- 106
HBU60- 107
HBU60- 108
HBU60- 109
HBU60- 110
Output Voltage
012 ~ 13 VDC
013 ~ 16 VDC
16 ~ 21 VDC
21 ~ 27 VDC
27 ~ 33 VDC
33 ~ 40 VDC
Output Current
5.00 ~ 4.61 A
4.61 ~ 3.75 A
3.75 ~ 2.85 A
2.85 ~ 2.22 A
2.22 ~ 1.81 A
1.81 ~ 1.50 A
Total Regulation
5%
5%
5%
3%
3%
3%
Maximum
Output Power
6 0W
6 0W
6 0W
6 0W
6 0W
6 0W
Mechanical specifications :
Pin Chart
PIN
MODEL
1
2
3
4
5
6
HBU60-1XX OUT OUT OUT RTN RTN RTN
Note:
1. Dimensions are shown in inches or mm.
2. Weight: 140gs approx.
3. Input connector mates with Molex housing
09-50-3031/35977-0300 and Molex 2478/ 35922
series crimp terminal.
4. Output connector mates with Molex housing
09-50-3061/35977-0600 and Molex 2478/ 35922
series crimp terminal.