English
Language : 

HBU60 Datasheet, PDF (1/2 Pages) MPS Industries, Inc. – 60W Open frame type medical power supplies for health care applications
HBU60 SERIES
60W Open frame type medical power supplies for health care applications
Description:
The HBU60 series of compact, open frame constructed, AC/DC
switching mode power supplies provide 60 Watts of continuous
output power .They are suited for use in health care applications.
All models meet FCC Part-18 class B and CISPR-11 EN55011 class
B emission limits and are designed to comply with UL/c-UL(UL
60601-1:2ndEdition) ,TUV/T-mark(EN 60601-1:2ndEdition) and new
CE requirements. All units are 100% burned in and tested.
Features:
Wide Operating voltage 90 to 264 VAC,47 to 63 Hz
Single output
Input connector mates with molex housing 09-50-3031/35977-0300
and molex 2478/ 35922 series crimp terminal.
Output connector mates with molex housing 09-50-3061/35977-0600
and molex 2478/ 35922 series crimp terminal.
Input surge current, over voltage and over load protection
Size: 2"x4”x1”
Class I
2 year warranty
Safety approvals :
Electrical characteristics:
CB
EN 60601-1
IEC 60601-1
Sym.
Parameter
Test Conditions
Min. Typ. Max. Unit
Vin
fin
Po
Vo
Io
Iil
Iih
I rl
I rh
Eff
REG-i
REG -o
OVP
OCP
Ttr
Thold
Ts
Vp-p
Ilk
TC
Safety Approvals Input
Voltage Range
100
240
Operate Voltage Range
90
264
Input Frequency
47
63
Output Power Range
Vin=90 to 264 VAC
0
60
Output Voltage Range
See rating Chart
Output Current Range
See rating Chart
Input Current (Low Line) Io=Full load, Vin=115VAC
1.62
Input Current (High Line) Io=Full load, Vin=230VAC
0.72
Low Line Inrush Current
High Line Inrush Current
Efficiency
Io=Full load, 25 C,Cool start,
Vin=115VAC
Io=Full load, 25 C,Cool start,
Vin=230VAC
Io=Full load, Vin=230VAC
26 30
43 47
85
Line Regulation
Io=Full load
1
Load Regulation
Vin=230VAC
5
Over Voltage Protection
Over Current Protection
112
132
110
150
Time of Transient Response Io=Full load to Half Load, Vin=100VAC
4
Hold-Up Time
Io=Full load, Vin=110VAC
12 16
Start Up Time
Io=Full load, Vin=100VAC
Ripple & Noise (Peak to Peak) Full load, Vin=90VAC
Safety Ground Leakage Current Io=Full load, Vin=240VAC
0.3 1 2
1
0.1 0.3
Temperature Coefficient All output
-0.04
0.04
VAC
VAC
Hz
W
V
A
A
A
A
A
%
%
%
%
%
mS
mS
S
%
mA
%/ C
Environmental :
Sym.
Parameter
T e s t Conditions
Toper Operating Temperature
Tstg Storage Temperature
Ho Operating Humidity
Hr Storage Humidity
MTBF Operating Temperature at 25 C, Calculated per MIL-HDBK-217F
Pd Derate linearly from 100% load at 40 C to 50% load at 70 C
Min. Typ. Max. Unit
0 40 70
C
-40
85
C
0
95 %
0
75 %
0.1M
Hrs