GENERAL CHARACTERISTICS
Power Adapter
Line voltage range 50/60 Hz, 100 – 240 VAC
Output voltage 12 VDC
Battery
Ni-MH rechargeable battery pack 7.2 V, 4500 mAH
Operating time: 4 hours
Charging time: 5 hours, measurement unit off
Allow ambient temperature during charging: 10 ºC to 40 ºC
Operating Environment
Temperature Operating full accuracy 0 ºC to 50 ºC
Non-operating
–20 ºC to 70 ºC
Humidity Operating full accuracy to 80% RH at 40 ºC
Altitude
Operating full accuracy Up to 2000 m
Non-operating
15000 m (50000 ft)
ESD tolerance ± 4 kV
Safety Compliance
IEC 61010-1: 2001/EN61010-1: 2001
CSA C22.2 No. 61010-1: 2004
Pollution degree 2
This instrument is rated for indoor use only.
Dimensions (HxWxD)
24.1 cm height x 13.8 cm width x 6.6 cm depth
Weight
1.5 kg
I/O
USB 2.0 Full Speed Client - For PC connectivity (Standard Option)
USB 2.0 Full Speed Host - For external USB storage (Option 001)
Firmware upgrade through USB.
Warranty
3 years for main unit
3 months for standard shipped accessories unless
otherwise stated.
ORDERING INFORMATION
Standard Shipped Items
Item
U1560A 1:1 Scope probe
U1561A 10:1 Scope probe
U1580A Test leads
U1162A Alligator clips
USB cable
U1570A AC power adapter and cord
U1571A Ni-MH battery pack
Quick Start Guide
Certiicate of calibration
U1602B/U1604B
-
x (2 units)
x
x
x
x
x
x
x
U1602B
U1604B