34465A Datasheet, PDF (8/23 Pages) Keysight Technologies – Digital Multimeters
08 | Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet
Specifications 34460A
Range 2/frequency
True RMS AC current 2, 6, 8
100 µA, 1 mA, 10 mA, and
100 mA ranges
3 Hz – 5 kHz
5 – 10 kHz (typ)
1 A range
3 Hz – 5 kHz
5 – 10 kHz (typ)
3 A range
3 Hz – 5 kHz
5 – 10 kHz (typ)
Continuity
1 kΩ
Diode test 9
5V
DC ratio (typ)
Burden voltage
< 0.011, < 0.11, < 0.05, < 0.5 V
< 0.7 V
< 2.0 V
Temperature 10
PT100 (DIN/ IEC 751)
5 kΩ thermistor
Frequency: specification ± (% of reading) 11, 12
100 mV, 1 V, 10 V, 100 V, and 750 V ranges 13
3 Hz – 10 Hz
10 Hz– 100 Hz
100 Hz – 1 kHz
1 Hz– 300 kHz
Square wave 14
Additional gate time errors ± ( % of reading ) 12, 10
Frequency
3 Hz– 40 Hz
40 Hz– 100 Hz
100 Hz – 1 kHz
1 Hz– 300 kHz
Square wave 14
24 hours 3
TCAL ± 1 °C
90 days
TCAL ± 5 °C
1 year
TCAL ± 5 °C
2 years
TCAL ± 5 °C
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.23 + 0.04
0.23 + 0.04
0.23 + 0.04
0.23 + 0.04
0.23 + 0.04
0.23 + 0.04
0.23 + 0.04
0.23 + 0.04
0.002 + 0.030 0.008 + 0.030 0.010 + 0.030 0.012 + 0.030
0.002 + 0.030 0.008 + 0.030 0.010 + 0.030 0.012 + 0.030
(normalized input accuracy) + (normalized reference accuracy)
Probe accuracy + 0.05 °C
Probe accuracy + 0.1 °C
0.100
0.030
0.030
0.002
0.001
1 second
0
0
0
0
0
0.100
0.030
0.010
0.008
0.008
0.1 second
0.200
0.060
0.020
0.004
0
0.100
0.030
0.012
0.012
0.012
0.100
0.035
0.017
0.017
0.017
0.01 second
0.200
0.200
0.200
0.030
0
Temperature
coefficient/°C 4
0.015 + 0.006
0.030 + 0.006
0.015 + 0.006
0.030 + 0.006
0.015 + 0.006
0.030 + 0.006
0.0010 + 0.0020
0.0010 + 0.0020
0.0002
0.0002
0.0002
0.0002
0.0002
1.  For DC: Specifications are for 60-minute warm-up, aperture of 10 or 100 NPLC, and auto zero on. For AC: Specifications are for 60-minute warm-up,
slow AC filter, sine wave.
2.  20% overrange on all ranges, except 1,000 V DCV, 750 ACV, 3 A AC, and diode test.
3.  Relative to calibration standards.
4.  Add this for each °C outside TCAL ± 5 °C.
5.  Specifications are for sine wave input > 0.3% of range and > 1 mVrms. 750 ACV range limited to 8 x 107 Volt–Hz.
6.  Low-frequency performance: three filter settings are available: 3 Hz, 20 Hz, 200 Hz. Frequencies greater than these filter settings are specified with no
additional errors.
7.  Specifications are for 4–wire ohms function or 2–wire ohms using math null for offset. Without math null, add 0.2 Ω additional error in 2-wire ohms function.
8.  Specifications are for sinewave input > 1% of range and > 10 µA AC.
9.  Specifications are for the voltage measured at the input terminals. The 1 mA test current is typical. Variation in the current source will create some
variation in the voltage drop across a diode junction.
10.  Actual measurement range and probe errors will be limited by the selected probe. Probe accuracy adder includes all measurement and ITS-90 temperature
conversion errors PT100 Ro settable to 100 Ω ± 5 Ω to remove the initial probe error.
11.  Specifications are for 60-minute warm-up and sine wave input unless stated otherwise. Specifications are for 1-second gate time (7 digits).
12.  Applies to sine and square inputs ≥ 100 mV. For 10 mV to < 100 mV inputs, multiply % of reading error x10.
13.  Amplitude 10% – 120% of range and less than 750 ACV.
14.  Square wave input specified for 10 – 300 kHz.
15.  Specifications are for using Math Null zeroing. High dissapation factor capacitors may show different results than a single frequency measurement.
Film capacitors usually have lower dissapation factors than other dielectrics.