English
Language : 

34465A Datasheet, PDF (21/23 Pages) Keysight Technologies – Digital Multimeters
21 | Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet
General Characteristics
(for all models except where noted)
Line power
Power supply
Power line frequency
Power consumption
Environment
Operating environment
Operating altitude
Storage temperature
Mechanical
Rack dimensions
Bench dimensions
Weight
Regulatory
Safety
EMC
Acoustic noise (nominal)
Triggering conditions
External input
Delay
Jitter
Minimum pulse width
Maximum rate
Voltmeter complete output
Polarity
Pulse width
Computer interfaces
LXI (rev 1.4)
USB
GPIB
Language
100/120 (127)/ 220 (230)/240 ACV ± 10%, CAT II
50/60/400 Hz ± 10%
25 VA
Full accuracy for 0 to 55 °C
Full accuracy to 80% R.H. at 40 °C non–condensing
Up to 3,000 m
–40 to 70 °C
(W x H x D): 212.8 mm x 88.3 mm x 272.3 mm
(W x H x D): 261.2 mm x 103.8 mm x 303.2 mm
34460A: 3.68 kg (8.1 lb)
34461/65/70A: 3.76 kg (8.3 lb)
EN 61010-1:2010 (3rd Edition)
ANSI/ISA-61010-1 (82.02.01) Third Edition
ANSI/UL 61010-1 Third Edition
CAN/CSA-C22.2 No. 61010-1 Third Edition
EN 61010-2-030:2010 (1st Edition)
ANSI/ISA-61010-2-030 (82.02.03) First Edition
ANSI/UL 61010-2-030 First Edition
CAN/CSA-C22.2 No. 61010-2-030 First Edition
Refer to Declaration of Conformity for current
revisions
Measurement Category II to 300 V
Other non MAINS circuits to 1,000 Vpk
Pollution Degree 2
IEC 61326
EN 61326
CISPR
ICES-001
AS/NZS 2064.1
Refer to Declaration of Conformity for current
revisions
35 dBA
Low–power TTL compatible input programmable
edge triggered
< 1 µs
< 1 µs
1 µs
Up to 1 kHz (34461A), up to 300 Hz (34460A)
3.3 V logic output
Programmable edge pulse
Approximately 2 µs
10/100Base-T Ethernet (Sockets, VXI-11 protocol,
Web user interface) (Optional on 34460A)
USB 2.0 (USB-TMC488 & MTP protocol)
Optional GPIB IEEE-488
SCPI-1999, IEEE-488.2, 34401A compatible
Front-panel USB host port
Supports USB 2.0 high-speed mass storage (MSC) class devices
Capability: Import/export instrument configuration files, save volatile
readings and screen captures
System speeds (nom)
Benchmark
GPIB
Function change 1 50/s
Range change 2
100/s
USB 2.0
50/s
100/s
VXI-11
50/s
100/s
Sockets
50/s
100/s
1.  Rate to change from 2-wire resistance to any other function
2.  Rate to change from one range to the next higher range, ≤ 10 V, ≤ 10 MΩ
Triggering and memory
Samples per trigger
1 to 1,000,000
Trigger delay
0 to 3600 sec (~1 µs step size)
External trigger delay < 10 µs
External trigger jitter < 1 µs (DC fixed range)
Volatile reading
memory
10,000 (34461A), 1,000 (34460A)
Probe hold
Capture and navigate stable list of readings
Internal flash file system
80 MB total capacity
Save reading memory to non-volatile memory in CSV format
Store and recall user-defined states, power-off state,1 and preference files
Save screen captures in BMP or PNG formats
1.  Power-off state only when power-down is initiated via front-panel power
switch.
Math functions
Per function null, min/max/avg/Sdev, dB, dBm, span, count, limit test,
histogram
Display
4.3” color TFT WQVGA (480x272) with LED backlight
Supports: basic number, bar meter, trend chart (34461A only), histogram
views. User-defined power-on message, display label, and selectable
screen colors Integrated, context-sensitive system help through
press-and-hold buttons
Real-time clock/calendar
Set and read, year, month, day, hour, minute, seconds (Note: seconds not
settable). Battery CR-2032 coin-type, replaceable, > 10-year life (typ)
Software available
IO Libraries: www.keysight.com/find/IOLibraries