English
Language : 

U1230 Datasheet, PDF (7/13 Pages) Agilent(Hewlett-Packard) – Handheld Digital Multimeter (DMM)
07 | Keysight | U1230 Series Handheld Digital Multimeters (DMMs) - Data Sheet
Electrical Specifications (continued)
AC specifications
True rms AC voltage and AC current specifications
Function
Voltage
Current1
Range
600 mV
6V
60 V
600 V
600 (VZLOW)3
60 μA2
600 μA2
6 A3,5
10 A3, 4
Resolution
0.1 mV
0.001 V
0.01 V
0.1 V
0.1 V
0.01 μA
0.1 μA
0.001 A
0.01 A
Accuracy ± (% of reading +
counts of least significant digit)
45 Hz to 500 Hz
500 Hz to 1 kHz
1.0% + 3
2.0% + 3
1.0% + 3
2.0% + 3
1.0% + 3
2.0% + 3
1.0% + 3
2.0% + 3
2.0% + 3
4.0% + 3
1.5% + 3
NA
1.5% + 3
NA
1.5% + 3
NA
1.5% + 3
NA
Burden voltage/shunt
Where (applicable)
NA
NA
NA
NA
NA
< 2.5 V/1 kΩ
< 2.5 V/1 kΩ
< 0.2 V/0.005 Ω
< 0.4 V/0.005 Ω
Notes for true rms ac voltage specifications:
1.  Overload protection: 600 Vrms. For millivolt measurements, 600 Vrms for short circuits with < 0.3 A current.
2.  Input impedance: 10 MΩ (nominal) in parallel with < 100 pF.
3.  VZLOW input impedance: 3 kΩ (nominal).
Notes for ac current specifications:
1.  AC current measurement not available for U1231A model.
2.  Overload protection for 60 μA to 600 μA range: 600 Vrms for short circuits with < 0.3 A current.
3.  Overload protection for 6 A to 10 A range: 11 A/1000 V; 10 × 38 mm fast-acting fuse.
4.  Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals > 10 A to 20 A for 30 seconds
maximum. After measuring currents > 10 A, cool down the multimeter for twice the duration of the measured time before proceeding with
low current measurements.
5.  AC current between 0.6 mA and 300 mA is not measureable on the U1232A and U1233A models.
Capacitance specifications
Range
1000 nF
10 μF
100 μF
1000 μF
10 mF
Resolution
1 nF
0.01 μF
0.1 μF
1 μF
0.01 mF
Accuracy± (% of reading +
counts of least significant digit)
U1231A/U1232A/U1233A
1.9% + 2
1.9% + 2
1.9% + 2
1.9% + 2
1.9% + 2
Measuring rate (at full scale)
4 times/second
4 times/second
4 times/second
1 time/second
0.1 time/second
1.  Overload protection: 600 Vrms for short circuits with < 0.3 A current.
2.  The accuracy of for all ranges is specified based on a film capacitor or better, and after the Null function is used to subtract the test lead
resistance and thermal effect (by opening the test leads).
3.  The maximum display is 1200 counts.