English
Language : 

N9916A Datasheet, PDF (7/48 Pages) Keysight Technologies – FieldFox Handheld Analyzers
07 | Keysight | FieldFox Handheld Analyzers 4/6.5/9/14/18/26.5/32/44/50 GHz - Data Sheet
Cable and Antenna Analyzer and Vector Network Analyzer (continued)
System performance specifications
System dynamic range 1,2 (dB), high power, 300 Hz IFBW, 100 point average, Port 1 or port 2 (-10 to 55 ºC)
Frequency
Spec
N991xA, N992xA
> 300 kHz to 9 GHz3
95
> 9 to 14 GHz
91
> 14 to 18 GHz
90
> 18 to 20 GHz
87
> 20 to 25 GHz
74
> 25 to 26.5 GHz
65
N995xA
> 300 kHz to 1 MHz
—
> 1 to 10 MHz
—
> 10 MHz to 20 GHz4
100
> 20 to 44 GHz5
90
> 44 to 50 GHz6
81
Temperature stability
Nominal
Frequency
Magnitude (dB/ºC)
N991xA, N992xA
≤ 15 GHz
± 0.018
> 15 to 26.5 GHz
± 0.080
N995xA
≤ 15 GHz
± 0.005
≤ 25 GHz
± 0.030
> 25 GHz
± 0.060
Measurement speed (Sweep time)
CAT
N991xA, N992xA
Return loss, 30 kHz to 26.5 GHz, 1-port cal, 1001 points7
433 μs /pt
Return loss, 300 kHz to 50 GHz, 1-port cal, 1001 points
—
Distance-to-fault, 100 meter cable, 1-port cal, 1001 points7
480 μs /pt
VNA
N991xA, N992xA
S11 and S21, 30 kHz to 26.5 GHz, enhanced response cal,
100 kHz IF bandwidth, 1001 points8
483 μs /pt
S11 and S21, 300 kHz to 50 GHz, enhanced response cal,
—
100 kHz IF bandwidth, 1001 points
Typical
100
97
94
90
79
70
70 (nominal)
100 (nominal)
110
100
90
Phase (deg/ºC)
—
—
± 0.1
± 0.3
± 0.6
N995xA
—
650 μs /pt
650 μs /pt
N995xA
—
580 μs /pt
1.  System dynamic range is measured in factory with loads on test ports after thru normalization, test port output power high.
2.  For CAT mode, “Insertion loss (2-port)”, decrease listed dynamic range specifications by 20 dB, as CAT mode IFBW is fixed at 10 kHz. Can obtain full
dynamic range by using S21 measurement in VNA mode with 100 Hz IFBW.
3.  < 300 kHz: 63 dB nominal; 2 to 9 MHz: 85 dB spec, 90 dB typical.
4.  Decrease by 3 dB between 15 to 15.8 GHz for S21.
5.  Decrease by 5 dB between 21.7 to 22.1 GHz for S21.
6.  Decrease by 4 dB between 44 to 50 GHz for S21.
7.  850 μs /pt; applicable for FieldFox with serial number prefix < than MY5607/SG5607/US5607 and FieldFox not upgraded with Option
N9910HU-100/200/300.
8.  850 μs /pt; applicable for FieldFox with serial number prefix < than MY5607/SG5607/US5607 and FieldFox not upgraded with Option
N9910HU-100/200/300/400.