06 | Keysight | U1730C Series Handheld LCR Meters – Data Sheet
U1731C/U1732C/U1733C Electrical Specifications
Inductance
Range
Resolution
20 μH
200 μH
2000 μH
20 mH
200 mH
2000 mH
20 H
200 H
2000 H
0.001 μH
0.01μH
0.1 μH
0.001 mH
0.01 mH
0.1 mH
0.001 H
0.01 H
0.1 H
100 Hz
NA
NA
0.7% + 10
0.5% + 3
0.5% + 3
0.2% + 3
0.2% + 3
0.7% + 5
1.0% + 5
Accuracy = AL + Offset
U1731C/U1732C/U1733C
U1732C/U1733C
120 Hz
1 kHz
10 kHz
NA
NA
1.0% + 5
NA
1.0% + 5
0.7% + 3
0.7% + 10
0.5% + 3
0.5% + 3
0.5% + 3
0.2% + 3
0.3% + 3
0.5% + 3
0.2% + 3
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 5
0.2% + 3
0.5% + 5
1.0% + 5
0.7% + 5
1.0% + 5
2.0% + 8
1.0% + 5
2.0% + 8
NA
U1733C
100 kHz
2.5% + 20
2.5% + 20
0.8% + 20
0.8% + 10
1.0% + 10
1.0% + 10
2.0% + 10
NA
NA
Phase Angle of Impedance
Range
Resolution
–180° ~180°
0.1°/1°
Accuracy (θe)
(AZ + Offset/Zx) x180/π
Condition
D < 1 or Q > 1
Impedance
Zx
AZ
1999.9 Ω
19999
0.2%
199.9 Ω
1999
0.2%
19.9 Ω
199
0.2%
1.9 Ω
19
0.2%
Offset
3
3
3
3
Notes:
a. Specifications are applicable to all models (U1731C, U1732C, and U1733C) unless specified
b. The “AZ” and Offset are the accuracy specified at impedance
c. The “π” is approximately 3.14159
θe
±0.12 °
±0.20 °
±0.98 °
±9.16 °
Dissipation/Quality Factor
Function
Range
Z
0.001~999
L
0.001~999
C
0.001~999
Accuracy (De)
AZ + Offset/Zx x 100% + 3
AL + Offset/Lx x 100% + 3
AC + Offset/Cx x 100% + 3
Condition
D < 1 or Q > 1
D < 1 or Q > 1
D < 1 or Q > 1
Capacitance
Cx
AC
88.88 μF
8888
0.2%
Offset
3
De
0.334% + 3
Notes:
1. Specifications are applicable to all models (U1731C, U1732C, and U1733C) unless specified
2. The “AZ, AL, AC” and Offset are the accuracy specified at Impedance, Inductance, and Capacitance, respectively
3. The Zx, Lx, and Cx are the display count of the reading. For example, the Cx is 8888 as if the capacitance is 88.88 μF for the range of
200 μF.
4. The Quality Factor is the reciprocal of Dissipation Factor