05 | Keysight | U1730C Series Handheld LCR Meters – Data Sheet
U1731C/U1732C/U1733C Electrical Specifications
Accuracy is given as ±(% of reading + counts of least significant digit) at 23 °C ± 5 °C, with relative humidity less than 80%. Please refer to
the User Guide about the measuring mode specified for each range of L/C/R, series or parallel mode. Measurements performed at the test
socket and necessary Open and Short corrections must prior be done. The accuracy is verified by design and specified type tests.
Impedance/Resistance
Accuracy = AZ + Offset
Range Resolution
U1731C/U1732C/U1733C
U1732C/U1733C
100 Hz
120 Hz
1 kHz
10 kHz
2 Ω1
0.0001 Ω
0.7% + 50
0.7% + 50
0.7% + 50
0.7% + 50
20 Ω1
0.001 Ω
0.7% + 8
0.7% + 8
0.7% + 8
0.7% + 8
200 Ω1
0.01 Ω
0.2% + 3
0.2% + 3
0.2% + 3
0.2% + 3
2000 Ω 0.1 Ω
0.2% + 3
0.2% + 3
0.2% + 3
0.2% + 3
20 kΩ
0.001 kΩ
0.2% + 3
0.2% + 3
0.2% + 3
0.2% + 3
200 kΩ 0.01 kΩ
0.5% + 5
0.5% + 5
0.5% + 5
0.5% + 5
2000 kΩ 0.1 kΩ
0.5% + 5
0.5% + 5
0.5% + 5
0.7% + 5
20 MΩ
0.001 MΩ
2.0% + 8
2.0% + 8
2.0% + 8
5.0% + 8
200 MΩ 0.01 MΩ
6.0% + 80
6.0% + 80
6.0% + 80
NA
The accuracy for ranges 2 Ω to 200 Ω is specified after Null function is used to subtract the resistance of test leads and the contact resistance.
U1733C
100 kHz
DCR
1.0% + 50
0.7% + 50
0.7% + 8
0.7% + 8
0.5% + 5
0.2% + 3
0.5% + 5
0.2% + 3
0.5% + 5
0.2% + 3
0.7% + 8
0.5% + 5
NA
0.5% + 5
NA
2.0% + 8
NA
6.0% + 80
Notes:
a. For the ranges of 20 MΩ and 200 MΩ, the R.H. is specified for < 60%
b. Resistance is specified to Q < 10 and D > 0.1, otherwise the accuracy is (AZ + Offset) x √ 1 + Q2
c. Equivalence Series Resistance (ESR) measurement is determined by impedance measurement and range. The maximum
display is up to 199.99 kΩ and the accuracy is (AZ + Offset) x √ 1 + Q2
Capacitance
Accuracy = AC + Offset
Range
Resolution
U1731C/U1732C/U1733C
U1732C/U1733C
100 Hz
120 Hz
1 kHz
10 kHz
20 mF
0.001 mF
0.5% + 8
0.5% + 8
NA
NA
2000 μF
0.1 μF
0.5% + 5
0.5% + 5
0.5% + 8
NA
200 μF
0.01 μF
0.3% + 3
0.3% + 3
0.5% + 5
0.5% + 8
20 μF
0.001 μF
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 5
2000 nF
0.1 nF
0.2% + 3
0.2% + 3
0.2% + 3
0.2% + 3
200 nF
0.01 nF
0.2% + 3
0.2% + 3
0.2% + 3
0.5% + 3
20 nF
0.001 nF
0.5% + 5
0.5% + 5
0.2% + 3
0.5% + 3
2000 pF1
0.1 pF
0.5% + 10
0.5% + 10
0.5% + 5
0.5% + 3
200 pF1
0.01 pF
NA
NA
0.5% + 10
0.8% + 10
20 pF1
0.001 pF
NA
NA
NA
1.0% + 20
This accuracy for the ranges of 20 pF~2000 pF is specified after Math Null which is used to substrate the stray capacitances for test leads.
U1733C
100 kHz
NA
NA
NA
5.0% + 10
0.7% + 10
0.7% + 10
0.7% + 10
2.0% + 10
2.0% + 10
2.5% + 10
Notes:
a. The accuracy for the ceramic capacitor will be influenced depending on the dielectric constant (K) of the material used to make
the ceramic capacitor. For related influence factors, please refer to the Component dependency factors section in the Impedance
Measurement Handbook, downloadable for free at http://www.keysight.com/find/lcrmeters