English
Language : 

34450A Datasheet, PDF (6/12 Pages) Keysight Technologies – 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM
6 | Keysight | 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM – Data Sheet
Temperature and capacitance specifications
Table 5. Temperature and capacitance accuracy ± (% of reading + % of range)
Function
Range1
Probe type or test current
Temperature
– 80.0 °C to 150 °C
– 110.0 °F to 300.0 °F
Capacitance
1.000 nF
10.00 nF
100.0 nF
1.000 μF
10.00 μF
100.0 μF
1.000 mF
10.00 mF
Notes on specifications:
1.  20% over range on all ranges.
5 kΩ thermistor probe
5 kΩ thermistor probe
100 nA
100 nA
1.0 μA
1.0 μA
10 μA
100 μA
0.5 mA
1.0 mA
Specifications are for 90 minutes warm-up time, slow
mode, and calibration temperature within 18 °C - 28 °C
1 year
23 °C ± 5 °C
Probe accuracy + 0.2 °C
Probe accuracy + 0.4 °F
—
1 + 0.5
1 + 0.5
1 + 0.5
1 + 0.5
1 + 0.5
1 + 0.5
2 + 0.5
Temperature coefficient
0 °C - 18 °C
28 °C - 55 °C
0.002 °C
0.0036 °F
—
0.02 + 0.001
0.02 + 0.001
0.02 + 0.001
0.02 + 0.001
0.02 + 0.001
0.02 + 0.001
0.02 + 0.001