English
Language : 

34450A Datasheet, PDF (5/12 Pages) Keysight Technologies – 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM
5 | Keysight | 34450A Multimeter 5.5 Digit Dual Display, Benchtop DMM – Data Sheet
AC specifications
Table 2. AC Accuracy ± (% of reading + % of range)
Function
Range1
True RMS AC voltage2
100.000 mV
Frequency
20 Hz – 45 Hz
Specifications are for 90 minutes warm-up time, slow
mode, and calibration temperature within 18 °C - 28 °C
1 year
23 °C ± 5 °C
1.0 + 0.1
Temperature coefficient
0 °C - 18 °C
28 °C - 55 °C
0.02 + 0.02
45 Hz – 10 kHz
0.2 + 0.1
0.02 + 0.02
10 kHz – 30 kHz
30 kHz – 100 kHz3
1.5 + 0.3
3.0 + 0.3
0.05 + 0.02
0.10 + 0.02
1.00000 V to 750.00 V
20 Hz – 45 Hz
1.0 + 0.14
0.02 + 0.02
45 Hz – 10 kHz
0.2 + 0.1
0.02 + 0.02
10 kHz – 30 kHz
1.5 + 0.3
0.05 + 0.02
True RMS AC current2
10.0000 mA to 10.0000 A
30 kHz – 100 kHz3
20 Hz – 45 Hz
3.0 + 0.35
1.5 + 0.1
0.10 + 0.02
0.02 + 0.02
45 Hz – 1 kHz
0.5 + 0.1
0.02 + 0.02
1 kHz – 10 kHz6
2.0 + 0.2
0.02 + 0.02
Notes on specifications:
1.  20% over range on all ranges except ACV 750 V and ACI 10 A
2.  Specifications are for sine-wave inputs more than the 5 % of range except 750 V range. Input signal must be more than 50 Vrms for 750 V range.
Maximum crest factor of 3 at full scale. Input impedance is 1 MΩ in parallel with capacitance less than 120 pF, AC couple with up to 400 DCV
3.  Additional error to be added as frequency > 30 kHz and signal input < 10% of range. 30 kHz to 100 kHz: 0.003% of full scale per kHz
4.  For input < 200 V rms
5.  For input < 300 V rms
6.  Frequencies > 5 kHz are typical for all ranges
Table 3. Frequency Accuracy ± (% of reading + 3 counts)
Function
Frequency
Range1
Frequency
100.000 mV to 750.00 V1
10.0000 mA to 10.0000 A
20 Hz – 300 kHz2
20 Hz – 10 kHz3
1 year
23 °C ± 5 °C
0.02 + 3
0.02 + 3
Temperature coefficient
0 °C - 18 °C
28 °C - 55 °C
0.005
0.005
Notes on specifications:
1.  The frequency can be measured up to 1 MHz as 0.5 V signal to 100 mV/1 V ranges.
2.  10% of range to full scale input on all ranges except where noted. 100 mV range specifications are for full scale or greater inputs.
For inputs from 10 mV to 100 mV, multiply total % of reading error by 10.
3.  10% of range to full scale input on all ranges except where noted. 10 mA range specifications are for full scale or greater inputs.
For inputs from 1 mA to 10 mA, multiply total % of reading error by 10.
Table 4. Frequency resolution
Function
Frequency
Range
100.000 mV to 750.00 V1
Frequency
119.999 Hz
1.19999 kHz
11.9999 kHz
119.999 kHz
1.19999 MHz
1.  The frequency can be measured up to 1 MHz as 0.5 V signal to 100 mV/1 V ranges.
Resolution
0.001 Hz
0.00001 kHz
0.0001 kHz
0.001 kHz
0.00001 MHz