English
Language : 

Z2090B Datasheet, PDF (5/7 Pages) Keysight Technologies – Modulation Domain and Time Interval Analyzers Pulse Analyzer System (PAS)
Keysight PAS System Characteristics Compared to Legacy HP Systems
System characteristics Keysight PAS system HP 5371A / 5372A
HP 53310A
Channels
2 channel TDC;
optional 4 channel digitizer
2 channel;
optional 3rd channel
2 channel;
optional 3rd channel
Frequency range
18 GHz
(TDC via IRM detector)
2 GHz on digitizer
500 MHz/2 GHz;
3rd channel optional;
external mixer/detector and
LO source for 2 to 18 GHz
200 MHz/2.5 GHz
3rd channel optional;
external mixer/detector and
LO source for 2 to 18 GHz
Bandwidth
80 MHz (TDC)
2 GHz (digitizer)
500 MHz/2 GHz optional
200 MHz/2.5 GHz optional
Time interval and resolution Unrestricted time capture
with 50 ps resolution
Up to 8 seconds with 150 ps Up to 1 second with 200 ps
resolution
resolution
Sustainable system
capture rate
1 M events/second
N/A;
accumulate statistics and
histogram data only
N/A;
accumulate statistics and
histogram data only
Data transfer rate
25M events/second
over PCI bus
25k measurements/second 25k measurements/second
over GPIB
over GPIB
System controller
Embedded compact PCI
controller
External via GPIB
External via GPIB
Measurement memory
4M events (TDC)
1 Gsample (digitizer)
8k measurements
8k measurements;
32k optional
Minimum PW
250 ps
1 ns
1 ns
Minimum PRI
15 ns
75 ns
75 ns
Measurement functions
PW, PRI, PRF vs time,
statistics, and histograms;
Raw frequency and time
interval data;
PW, PRI, PRF vs time,
statistics, histograms, duty
cycle;
Jitter and Allan variance
Extensive vector modulation
analysis with VSA option
Raw frequency and time
interval data (very limited
capture time)
PW, PRI, PRF vs time,
statistics, histograms, duty
cycle jitter and Allan variance
Raw frequency and time
interval data (very limited
capture time)
Basic modulation analysis
Basic modulation analysis
Triggering functions
Internal and external
Internal and external;
some advanced arming
functions to optimize use of
limited time capture
Internal and external;
some advanced arming
functions to optimize use of
limited time capture
5