English
Language : 

N2807B Datasheet, PDF (4/14 Pages) Keysight Technologies – N2807A PrecisionProbe Advanced Kit
Figure 4. Eye diagram with no correction,
notice the minimized eye margins due to
fixture loss
Figure 5. Same setup using PrecisionCable
correction to compensate for fixture loss
The software quickly (less than five minutes in most cases) and accurately
characterizes the desired element in the system without adding more equipment.
PrecisionProbe
– Characterizes probe input impedance
– Properly Creates Custom Probe Transfer Function = VOut / VIn or VOut/VInc =
VOut/VSrc
– Removes unwanted cable loss
Now every probe and cable in the system can have the exact same response
probe to probe or cable to cable, without the inaccuraies that using one model
can produce. Custom probes can now be properly characterized and unwanted
responses can be removed.
Not only does PrecisionProbe characterize the cables, it allows for immediate
use on the same instrument. When combining PrecisionProbe with the 90000
Q-Series, 90000 X-Series, and 90000A Series oscilloscopes, one can characterize
their measuring system and be using it within five minutes without adding more
complicated, expensive equipment. PrecisionProbe saves time and money while
increasing accuracy.
When combining InfiniiMax probes with switches between the amplifier and the
probe head, PrecisionProbe allows for full correction and automation of each
probe’s path. Full automation is then available to allow for quick swapping of the
inputs.
4