Four-Wire Resistance
Accuracy Conditions: Auto zero on, one hour warmup. Temperature within ±5°C of calibration temperature (module calibrated at 18-28°C).
Range
256 Ω
2 kΩ
16 kΩ
131 kΩ
1 MΩ
Source
Current
488 µA
488 µA
61 µA
61 µA
7.6 µA
Maximum Open
Circuit Voltage
11.5 V
11.5 V
11.5 V
11.5 V
11.5 V
RESOLUTION vs APERTURE
(Ω)
20/16.7 ms
10 µs
250 µΩ
2 mΩ
15 mΩ
125 mΩ
1Ω
15 mΩ
125 mΩ
1Ω
8Ω
64 Ω
90-DAY ACCURACY vs APERTURE
± (% of Reading + Ω)
20/16.7 ms
10 µs
0.035% + 10 mΩ
0.025% + 20 mΩ
0.025% + 200 mΩ
0.025% + 1 Ω
0.015% + 10 Ω
0.12% + 50 mΩ
0.1% + 200 mΩ
0.1% + 2 Ω
0.1% + 16 Ω
0.1% + 120 Ω
NOTE: With offset compensation on, accuracy is the same as for the voltmeter alone.
True RMS ac Voltage (ac coupled)
Crest Factor: 7 at 0% full scale; 5 at full scale. Accuracy Conditions: Sine wave inputs >0% of full scale. dc component <0% of ac component.
Auto-zero on, one hour warmup. Temperature within ± 5°C of calibration temperature (module calibrated at 8-28°C).
Range
(RMS)
87.5 mV
700 mV
Input
Impedance
>100 MΩ,
<100 pF
>100 MΩ,
<100 pF
5.6 V
>100 MΩ,
<100 pF
44.8 V
10 MΩ ± 5%,
<100 pF
300 V
10 MΩ ± 5%,
<100 pF
Frequency
20-50 Hz
50 Hz-1 kHz
5-10 kHz
20-50 Hz
50 Hz-1 kHz
1-5 kHz
5-10 kHz
20-50 Hz
50 Hz-1 kHz
1-5 kHz
5-10 kHz
20-50 Hz
50 Hz-1 kHz
1-5 kHz
5-10 kHz
20-50 Hz
50 Hz-1 kHz
1-5 kHz
5-10 kHz
RESOLUTION vs APERTURE
(Volts)
320/267 ms
10 µs
30 nV
7.6 µV
0.24 µV
61 µV
2.0 µV
488 µV
15 µV
3.9 mV
122 µV
31 mV
90-DAY ACCURACY vs APERTURE
± (% of Reading + Volts)
320/267 ms All other apertures
2.175% +200 µV 2.175% +1 mV
0.675% +200 µV 0.675% +200 µV
3.175% +200 µV 3.175% +200 µV
2.125% +1.5 mV 2.125% +8 mV
0.625% +1.5 mV 0.625% +1.5 mV
0.625% +1.5 mV 0.625% +1.5 mV
3.125% +1.5 mV 3.125% +1.5 mV
2.125% +15 mV 2.125% +80 mV
0.625% +15 mV 0.625% +15 mV
1.125% +15 mV 1.125% +15 mV
10.125% +15 mV 10.125% +15 mV
2.125% +100 mV 2.125% + 500 mV
0.625% +100 mV 0.625% +100 mV
1.125% +100 mV 1.125% +100 mV
10.125% +100 mV 10.125% +100 mV
2.125% + 500 mV 2.125% + 2.5 V
0.625% + 500 mV 0.625% + 500 mV
1.125% + 500 mV 1.125% + 500 mV
10.125% + 500 mV 10.125% + 500 mV
Timing/Synchronization
Timer/pacer
Timer range:
Resolution:
Programmable delay
Delay range:
Resolution:
External trigger
Minimum pulse width:
Maximum trigger rate:
Isolation
76 µs to 65.5 ms
2 µs
40 µs to 16 s
2 µs
100 ns
5 kHz (Trigger Condition, negative
edge; Fixed range,10 µs aperture)
450 Vpk between any terminal
and chassis.
dc Voltage Accuracy with Relay Multiplexers
Range 20/16.7 ms
10 µs
20/16.7 ms
10 µs
125 mV 0.023%+9 µV 0.115%+64 µV 0.023%+55 µV 0.115%+110 µV
1 V 0.013%+19 µV 0.1%+204 µV 0.013%+65 µV 0.1%+250 µV
8 V 0.01%+54 µV 0.1%+1.5 mV 0.01%+100 µV 0.1%+1.55 mV
64 V 0.015%+1 mV 0.1%+20 mV 0.015%+1.05 mV 0.1%+20 mV
300 V 0.015%+5 mV 0.1%+80 mV 0.015%+5.05 mV 0.1%+80 mV
Accuracy Conditions: Auto zero on, one hour warmup.
Temperature within ±5°C of calibration temperature
(module calibrated at 8-28°C).