Reading Rate
Auto zero off, fixed range, default trigger delay, offset comp off,
Sample Source “TIMER” for rates >15 readings/s.
Maximum reading rate: 13 K
Typical Reading Rates (rdgs/s)
APERTURE
320 ms 267 ms 20 ms 16.7 ms 2.5 ms 100 µs 10 µs
dc voltage: 3 3.5 49 59 365 3125 13000
Four-wire
resistance: 3 3.5 49 59 365 3125 13000
ac voltage: 1.3 1.4 1.9 1.9 1.9 1.9 1.9
Resolution (bits/digit)
APERTURE
320 ms 267 ms 20 ms 16.7 ms 2.5 ms 100 µs 10 µs
Binary bits: ±22 ±22 ±20 ±20 ±18 ±15 ±14
Decimal
digits:
6.5 6.5 6
6
5.5 4.5 4
Noise Rejection (dB) for DC voltage and Resistance functions.
Noise Rejection Conditions: CMR measured with 1 kΩ in both High and Low leads with a 10% imbalance. Low connted to Common at source,
measured with respect to earth ground. NMR is for specified frequencies ±0.1%
320 ms 267 ms
20 ms
APERTURE
16.7 ms
2.5 ms
100 µs
10 µs
dc:
Common mode rejection
ac (0-400): Common mode rejection*
150 dB
70 dB
150 dB
70 dB
150 dB
70 dB
150 dB
70 dB
150 dB
70 dB
150 dB
70 dB
150 dB
70 dB
50 Hz:
Power line cycles (NPLC)
16
Normal mode rejection
84 dB
Effective Common Mode ECMR* 154 dB
–
0 dB
70 dB
1
60 dB
130 dB
–
0 dB
70 dB
–
0 dB
70 dB
–
0 dB
70 dB
–
0 dB
70 dB
60 Hz:
Power line cycles (NPLCs)
–
Normal mode (60 Hz) rejection 0 dB
Effective Common Mode Rejection* 70 dB
16
84 dB
154 dB
–
0 dB
70 dB
1
60 dB
130 dB
–
0 dB
70 dB
–
0 dB
70 dB
–
0 dB
70 dB
400 Hz: Power line cycles (NPLCs)
Normal mode (400 Hz) rejection
Effective Common Mode ECMR*
* 64 & 300 volt ranges reduced by 36 dB
128
84 dB
154 dB
–
0 dB
70 dB
8
84 dB
154 dB
–
0 dB
70 dB
1
60 dB
130 dB
–
0 dB
70 dB
–
0 dB
70 dB
dc Voltage
Accuracy Conditions: Auto zero on,one hour warmup. Temperature within ±5°C of calibration temperature (module calibrated at 18-28°C).
Range
125 mV
1V
8V
64 V
300 V
Input Resistance
>100 MΩ
>100 MΩ
>100 MΩ
10 MΩ + 5%
10 MΩ + 5%
RESOLUTION vs APERTURE
(Volts)
20/16.7 ms
10 µs
120 nV
1.0 µV
7.6 µV
61 µV
488 µV
7.6 µV
61 µV
488 µV
3.9 mV
31 mV
90-DAY ACCURACY vs APERTURE
± (% of Reading + Volts)
20/16.7 ms
10 µs
0.023% + 5 µV
0.013% + 15 µV
0.01% + 50 µV
0.015% + 1 mV
0.015% + 5 mV
0.115% + 60 µV
0.1% + 200 µV
0.1% + 1.5 µV