W6155A Datasheet, PDF (10/16 Pages) Keysight Technologies – X-Series Measurement Application | |||
|
◁ |
10 | Keysight | ISDB-T/Tmm X-Series Measurement Application N6155A & W6155A - Technical Overview
Key Specifications
Definitions
ââ Specifications describe the performance of parameters covered by the
product warranty.
ââ 95th percentile values indicate the breadth of the population (â2Ï) of
performance tolerances expected to be met in 95% of cases with a 95%
confidence. These values are not covered by the product warranty.
ââ Typical values are designated with the abbreviation âtyp.â These are per-
formance beyond specification that 80% of the units exhibit with a 95%
confidence. These values are not covered by the product warranty.
ââ Nominal values are designated with the abbreviation ânom.â These values
indicate expected performance, or describe product performance that is
useful in the application of the product, but is not covered by the product
warranty.
ââ PXA specifications apply to analyzers with frequency options of 526 and
lower. For analyzers with higher frequency options, specifications are not
warranted but performance will nominally be close to that shown in this
section.
Note: Data subject to change
You Can Upgrade!
Options can be added after your
initial purchase.
All of our X-Series
application options
are license-key
upgradeable.
UP
GRADE
Description
PXA
Channel power
5.60 MHz integration bandwidth â50 dBm (nom)
Absolute power accuracy
20 to 30 °C
± 0.61 dB
(± 0.19 dB 95%)
Measurement floor
â87.2 dBm
Channel power with shoulder attenuation view
5.60 MHz integration bandwidth
ML1 = â14.0 dBm (nom)
Dynamic range, relative
Offset frequency 3.4 MHz
92.9 dB
(98.7 dB typ)
Power statistics CCDF
Minimum power at RF input
â50 dBm (nom)
Histogram resolution
0.01 dB
Adjacent channel power
Minimum power at RF input;
0 to 55 °C
â36 dBm (nom)
ACPR accuracy
Offset frequency 6 MHz
± 0.16 dB
1.â ML (mixer level) is RF input power minus attenuation
MXA
EXA
â50 dBm (nom)
± 0.82 dB
(± 0.23 dB 95%)
â84.2 dBm
â50 dBm (nom)
± 0.94 dB
(± 0.27 dB 95%)
â80.2 dBm
ML1 = â16.0 dBm (nom) ML1 = â16.0 dBm (nom)
87.7 dB
(94.1 dB typ)
82.5 dB
(89.6 dB typ)
â50 dBm (nom)
0.01 dB
â50 dBm (nom)
0.01 dB
â36 dBm (nom)
â36 dBm (nom)
5.60 MHz noise bandwidth, method = IBW
± 0.38 dB
± 0.81 dB
CXA
â50 dBm (nom)
± 1.33 dB
(± 0.61 dB 95%)
â77.2 dBm
ML1 = â15.0 dBm (nom)
78.9 dB
(86.6 dB typ)
â50 dBm (nom)
0.01 dB
â36 dBm (nom)
± 1.20 dB
|
▷ |