N1419A Datasheet, PDF (10/16 Pages) Keysight Technologies – Femto/Picoammeter and Electrometer/High Resistance Meter
10 | Keysight | B2980A Series Femto/Picoammeter and Electrometer/High Resistance Meter - Data Sheet
Technical Specifications and Characteristics
Specification conditions
Temperature: 23 °C ± 5 °C
Humidity: 30% to 80% RH
After 1 hour warm-up
Ambient temperature change: less than ± 3 °C after self-calibration
Calibration period: 1 year
Current measurement
Measurement range
Display resolution
Accuracy
± (% + offset)
2 pA
1 aA
1 + 3 fA
20 pA
10 aA
0.5 + 3 fA
200 pA
100 aA
0.5 + 5 fA
2 nA
1 fA
0.2 + 300 fA
20 nA
10 fA
0.2 + 500 fA
200 nA
100 fA
0.2 + 5 pA
2 μA
1 pA
0.1 + 50 pA
20 μA
10 pA
0.05 + 500 pA
200 μA
100 pA
0.05 + 5 nA
2 mA
1 nA
0.05 + 50 nA
20 mA
10 nA
0.05 + 500 nA
Temperature coefficient 0 to 18 °C and 28 to 45 °C
± (0.05 x Accuracy)/°C
Input burden voltage at
SelfCal ± 3 °C1
20 μV
20 μV
20 μV
20 μV
20 μV
20 μV
20 μV
20 μV
100 μV
1 mV
6 mV
Supplemental characteristics
Temperature coefficient of input voltage burden
RMS noise
NMRR3
Maximum input capacitance
< 10 μV/°C on pA, nA and μA ranges
140 aA for 2 pA range, 10 s duration, no cable, open cap
> 60 dB
10 nF on less than 20 μA ranges, 1 μF on other ranges
Notes
Conditions: properly zeroed, 6½-digit, 1 PLC, median filter on, moving average 10 points
1.  Supplemental characteristics
2.  0.1% of final value, step size 0% to 100% of range
3.  Normal mode rejection ratio, integration time = 1, 2, …, 100 PLC; power line frequency ± 0.1%
Measurement settling
time1, 2
16 s
1.4 s
1.4 s
13 ms
13 ms
1.2 ms
550 μs
600 μs
600 μs
100 μs
100 μs
Resistance measurement
Measurement range
Display resolution
1 MΩ
1Ω
10 MΩ
10 Ω
100 MΩ
100 Ω
1 GΩ
1 kΩ
10 GΩ
10 kΩ
100 GΩ
100 kΩ
1 TΩ
1 MΩ
10 TΩ
10 MΩ
100 TΩ
100 MΩ
1 PΩ
1 GΩ
Temperature coefficient 0 to 18 °C and 28 to 45 °C
Accuracy ± (% + offset)1, 2
0.135 + 1 Ω
0.135 + 10 Ω
0.185 + 100 Ω
0.285 + 1 kΩ
0.285 + 10 kΩ
0.41 + 100 kΩ
0.45 + 1 MΩ
0.625 + 10 MΩ
0.75 + 100 MΩ
2.6 + 1 GΩ
± (0.1 x Accuracy)/°C
Auto voltage source
20 V
20 V
20 V
20 V
20 V
20 V
200 V
200 V
200 V
200 V
Current measure range
200 μA
20 μA
2 μA
200 nA
20 nA
2 nA
2 nA
200 pA
20 pA
2 pA
Notes
Conditions: Auto V-source ohms, properly zeroed, 6½-Digit, 1 PLC, median filter on, digital filter = 10 readings.
1.  In the manual mode resistance can be calculated from specific source voltage and measured current. The measurement accuracy in the manual mode is
determined by voltage source accuracy and ammeter accuracy as follows: Measurement Error = R reading x (Voltage % error + Voltage offset error/
Voltage + Current measurement % error + 10 x Current measurement offset error / Current measurement range)
2.  Current measurement range for both Auto and Manual modes: 10% of current range ≤ measured current ≤ 100% of current range