34461A Datasheet, PDF (10/23 Pages) Keysight Technologies – Digital Multimeters
10 | Keysight | Digital Multimeters: 34460/61/65/70A - Data Sheet
Specifications 34461A
Range 2/frequency
True RMS AC current 2,6,9
100 µA, 1 mA, 10 mA,
and 100 mA ranges
3 Hz – 5 kHz
5 – 10 kHz (typ)
1 A range
3 Hz – 5 kHz
5 – 10 kHz (typ)
3 A range
3 Hz – 5 kHz
5 – 10 kHz (typ)
10 A range 8
3 Hz – 5 kHz
5 – 10 kHz (typ)
Continuity
1 kΩ
Diode test 10
5V
DC ratio (typ)
Burden voltage
< 0.011, < 0.11, < 0.05, < 0.5 V
< 0.7 V
< 2.0 V
< 0.5 V
Temperature 11
PT100 (DIN/ IEC 751)
5 kΩ thermistor
Frequency: specification ± (% of reading) 12, 13
100 mV, 1 V, 10 V, 100 V, and 750 V ranges 14
3 – 10 Hz
10 – 100 Hz
100 Hz – 1 kHz
1 – 300 kHz
Square wave 15
Additional gate time errors ± (% of reading) 13
Frequency
3 Hz – 40 Hz
40 Hz– 100 Hz
100 Hz – 1 kHz
1 kHz– 300 kHz
Square wave 15
24 hours 3
TCAL ± 1 °C
90 days
TCAL ± 5 °C
1 year
TCAL ± 5 °C
2 years
TCAL ± 5 °C
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.10 + 0.04
0.23 + 0.04
0.23 + 0.04
0.23 + 0.04
0.23 + 0.04
0.23 + 0.04
0.23 + 0.04
0.23 + 0.04
0.23 + 0.04
0.15 + 0.04
0.15 + 0.04
0.15 + 0.04
0.15 + 0.04
0.15 + 0.04
0.15 + 0.04
0.15 + 0.04
0.15 + 0.04
0.002 + 0.030 0.008 + 0.030 0.010 + 0.030 0.012 + 0.030
0.002 + 0.030 0.008 + 0.030 0.010 + 0.030 0.012 + 0.030
(normalized input accuracy) + (normalized reference accuracy)
Probe accuracy + 0.05 °C
Probe accuracy + 0.1 °C
0.100
0.030
0.003
0.002
0.001
1 second
0
0
0
0
0
0.100
0.030
0.008
0.006
0.006
0.1 second
0.200
0.060
0.020
0.004
0
0.100
0.030
0.010
0.010
0.010
0.100
0.030
0.010
0.010
0.010
0.01 second
0.200
0.200
0.200
0.030
0
Temperature
coefficient/°C 4
0.015 + 0.006
0.030 + 0.006
0.015 + 0.006
0.030 + 0.006
0.015 + 0.006
0.030 + 0.006
0.015 + 0.006
0.030 + 0.006
0.0010 + 0.0020
0.0010 + 0.0020
0.100
0.035
0.015
0.015
0.015
1.  For DC: Specifications are for 60-minute warm-up, aperture of 10 or 100 NPLC, and auto zero on. For AC: Specifications are for 60-minute warm-up, slow
AC filter, sine wave.
2.  20% over range on all ranges, except 1,000 V DCV, 750 ACV, 10 A DC, 3 A AC, 10 A AC, and diode test.
3.  Relative to calibration standards.
4.  Add this for each °C outside TCAL ± 5 °C.
5.  Specifications are for sinewave input >0.3% of range and > 1 mVrms. 750 ACV range limited to 8 x 10 7 Volt–Hz.
6.  Low-frequency performance: three filter settings are available: 3 Hz, 20 Hz, 200 Hz. Frequencies greater than these filter settings are specified with no
additional errors.
7.  Specifications are for 4-wire ohms function or 2-wire ohms using math null for offset. Without math null, add 0.2 Ω additional error in 2-wire ohms function.
8.  The 10 A range is only available on a separate front-panel connector. Add 2 mA base per amp or inputs > 5 A rms.
9.  Specifications are for sinewave input > 1% of range and > 10 µA AC.
10.  Specifications are for the voltage measured at the input terminals. The 1 mA test current is typical. Variation in the current source will create some
variation in the voltage drop across a diode junction.
11.  Actual measurement range and probe errors will be limited by the selected probe. Probe accuracy adder includes all measurement and ITS-90
temperature conversion errors. PT100 Ro settable to 100 Ω ± 5 Ω to remove the initial probe error.
12.  Specifications are for 60-minute warm-up and sine wave input unless stated otherwise. Specifications are for 1-second gate time (7-digits).
13.  Applies to sine and square inputs ≥ 100 mV. For 10 mV to < 100 mV inputs, multiply % of reading error x10.
14.  Amplitude 10% – 120% of range and less than 750 ACV.
15.  Square wave input specified for 10 – 300 kHz.