Performance Characteristics (continued)
Vertical system: oscilloscope channels (continued)
Single-cursor accuracy
Dual-cursor accuracy
±{DC vertical gain accuracy + DC vertical offset accuracy + 0.2% full scale (~1/2 LSB)}
Example: for 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
accuracy = ±{4.0% (80 mV) + 0.1(10 mV) + 2.0 mV + 0.5% (5 mV) + 0.2% (80 mV)}
= ±6.385 mV
±{DC vertical gain accuracy + 0.4% full scale (~1 LSB)}
Example: for 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
accuracy = ±{4.0% (80 mV) + 0.4% (80 mV)} = ±3.52 mV
Horizontal
Range
1 ns/div to 50 s/div
Time base accuracy
20 ppm
Delay range
Pre-trigger: –100 %
Post-trigger: +100 %
Modes
Main, roll, XY
XY
Yes
Reference position
Center
Trigger system
Sources
Modes
Ch 1, Ch 2, Ext (not applicable for TV trigger)
Normal, single, auto trigger
Holdoff time
Selections
60 ns
Edge, pulse width, TV
Edge
Triggers on a rising or falling edge, alternating, or either edge of any source
Pulse width
TV
Triggers on a pulse width greater than, equal to, or less than a specified time limit, with
time limits ranging from 16 ns to 10 s.
Minimum lower limit: 8 ns
Minimum upper limit: 16 ns
Maximum pulse width setting: 10 s
Triggers on one of three standard television waveforms: NTSC, PAL, SECAM
TV trigger sensitivity: 0.6 division of sync signal. Modes supported include Field 1, Field
2, all fields, or any line within a field.
AutoScale
Single-button automatic setup of all channels
Oscilloscope channel triggering
Range (internal)
±4 div from center screen
Coupling
AC (< 15 Hz)
LF reject (~ 35 kHz)
HF reject (~ 35 kHz)
External (EXT) triggering
Input impedance
Maximum input
Range
EXT trigger pulse width
1 MΩ: _ 16 pF
CAT I 30 Vrms, 42 Vpk
DC coupling: trigger level ±1.25 V and ±2.5 V
> 2.5 ns
9