U2701A Datasheet, PDF (8/11 Pages) Agilent(Hewlett-Packard) – U2701A and U2702A USB Modular Oscilloscope
Performance Characteristics[1]
Acquisition: oscilloscope channels
Real time sample rate
2 channels interleaved
Each channel
1 GSa/s
500 MSa/s
Standard memory depth
2 channels interleaved
Each channel
Normal
32 Mpts
16 Mpts
Single-shot
64 Mpts
32 Mpts
Vertical resolution
8 bits
Peak detection
Yes
Averaging
Any number from 1 to 999
Filter
Sin(x)/x interpolation for time base 1 ns to 100 ns
Sweep modes
Auto, normal, single
Vertical system: oscilloscope channels
Scope channels
U2701A/U2702A: Ch 1 and 2 simultaneous acquisition
AC coupled
U2701A: 3.5 Hz to 100 MHz
U2702A: 3.5 Hz to 200 MHz
Calculated rise time
(= 0.35/bandwidth)
U2701A: 3.5 ns
U2702A: 1.75 ns
Single-shot bandwidth
Range
Maximum input[2]
Offset range
Dynamic range
U2701A: 100 MHz
U2702A: 200 MHz
2 mV/div to 5 V/div (1 MΩ)
CAT I 30 Vrms, 42 Vpk
±4 div
Example: ±8 mV on 2 mV/div; ±20 V on 5 V/div
±4 div
Input impedance
Coupling
BW limit
Standard probes
ESD tolerance
Noise peak-to-peak
DC vertical offset accuracy
DC vertical gain accuracy
1 MΩ: _ 16 pF
AC, DC, Ground
_ 25 MHz
10:1 Passive Probe 150 MHz 1.2 m
10:1 Passive Probe 300 MHz 1.2 m
±2 kV
3 mVpp
≤ 200 mV/div: ±0.1 div ±2.0 mV ±0.5% offset value;
> 200 mV/div: ±0.1 div ±2.0 mV ±1.5% offset value
±4.0% of Full Scale
1 All characteristics are typical performance values and are not warranted. Characteristics are valid after a 30-minute warm-up
period and within ±10ºC of last calibration temperature.
2 Under standalone use, users are only allowed to measure up to CAT1 30 Vrms. For high-voltage measurement up to
CAT1 300 Vrms, users must install the L-mount kit on the U2701A/U2702A before plugging into the instrument chassis. Ensure
that the L-Mount kit installed on your modular oscilloscope is screwed to the instrument chassis to ensure chassis grounding. It is
required to use the provided 10:1 probes (N2862A/N2863A) for high-voltage measurements