U1253B Datasheet, PDF (7/14 Pages) Agilent(Hewlett-Packard) – Agilent U1250 Series Handheld Digital Multimeters
FREQUENCY SPECIFICATIONS[4]
RANGE
99.999 Hz
999.99 Hz
9.9999 kHz
99.999 kHz
999.99 kHz
RESOLUTION
0.001 Hz
0.01 Hz
0.0001 kHz
0.001 kHz
0.01 kHz
ACCURACY
U1251A/2A
U1253A
0.02%+3[8]
0.02%+3[8]
0.02%+3, <600 kHz 0.02%+3, <600 kHz
MIN. INPUT
FREQUENCY
1 Hz
[1] Input impedance: Refer to Table A on page 8.
[2] Crest factor ≤ 3.0 at full scale, 5.0 at half scale except the 1000 mV and 1000 V ranges where it is 1.5 at full scale, 3.0 at half scale.
For non-sinusoidal waveform, add 0.1% of reading ± 0.3% of range.
[3] The additional error to be added as frequency >20 kHz and signal input <10% of range: 3 counts of LSD per kHz.
[4] The input signal is lower than 20,000,000 V-Hz (the product of voltage and frequency).
[5] Input current >35 µArms.
[6] Current can be measured from 2.5 A up to 10 A continuously. An additional 0.5% needs to be added to the specified accuracy if the signal measured is in the range
of 10 A ~ 20 A for 30 seconds maximum. After measuring a current of >10 A, leave the meter to cool down for twice the measuring time used before application of
low current measurement.
[7] Input current < 3 Arms.
[8] For non-square wave signals, add 5 counts.
U1251A/U1251B FREQUENCY SENSITIVITY DURING VOLTAGE MEASUREMENT
INPUT RANGE
(Maximum input for specified
accuracy = 10 x Range or 1000 V)
50.000 mV
500.00 mV
1000.0 mV
5.0000 V
50.000 V
500.00 V
1000.0 V
FREQUENCY SENSITIVITY AND TRIGGER LEVEL
MINIMUM SENSITIVITY
(R.M.S. Sine Wave)
TRIGGER LEVEL FOR DC COUPLING
20 Hz - 100 kHz
>100 kHz ~ 200 kHz
< 100 kHz
>100 kHz ~ 200 kHz
10 mV
25 mV
40 mV
0.25 V
2.5 V
25 V
50 V
15 mV
35 mV
50 mV
0.5 V
5V
N/A
N/A
10 mV
60 mV
100 mV
0.5 V / 1.25 V (< 100 Hz)
5V
50 V
300 V
15 mV
70 mV
150 mV
0.6 V
6V
N/A
N/A
U1252A/U1252B/U1253A/U1253B FREQUENCY SENSITIVITY DURING VOLTAGE MEASUREMENT
FREQUENCY SENSITIVITY AND TRIGGER LEVEL
INPUT RANGE
MINIMUM SENSITIVITY
(R.M.S. Sine Wave)
TRIGGER LEVEL FOR DC COUPLING
(Maximum input for specified
accuracy = 10 x Range or 1000 V)
20 Hz ~ 200 kHz
>200 kHz ~ 500 kHz
< 100 kHz
>100 kHz ~ 500 kHz
50.000 mV
10 mV
25 mV
10 mV
25 mV
500.00 mV
70 mV
150 mV
70 mV
150 mV
1000.0 mV
120 mV
300 mV
120 mV
300 mV
5.0000 V
0.3 V
1.2 V
0.6 V
1.5 V
50.000 V
3V
5V
6V
15 V
500.00 V
30 V < 100 kHz
N/A
60 V
N/A
1000.0 V
50 V < 100 kHz
N/A
120 V
N/A
6