English
Language : 

U1253B Datasheet, PDF (6/14 Pages) Agilent(Hewlett-Packard) – Agilent U1250 Series Handheld Digital Multimeters
U1252A/U1252B/U1253A/U1253B AC+DC SPECIFICATIONS
FUNCTION
TRUE RMS AC
VOLTAGE[1] [2]
RANGE
50.000 mV
500.00 mV
1000.0 mV
5.0000 V
50.000 V
500.00 V
1000.00 V
RESOLUTION
0.001 mV
0.01 mV
0.1 mV
0.0001 V
0.001 V
0.01 V
0.1 V
ACCURACY ± (% of reading + No. of Least Significant Digit)
FREQUENCY
30 Hz ~ 45 Hz 45 Hz ~ 1 kHz 1 kHz ~ 10 kHz 10 kHz ~ 20 kHz 20 kHz~100 kHz[3]
1.5+80
0.4+60
0.7+60
0.8+60
3.5+220
1.5+65
0.4+30
0.4+45
0.4+30
0.6+30
0.4+30
0.4+45
0.8+45
1.5+45
1.5+45[4]
3.5+125
3.5+125[4]
N/A
FUNCTION
RANGE
RESOLUTION
AC CURRENT
500.00 µA[5]
5000.0 µA
50.000 mA
440.00 mA
5.0000 A
10.000 A[6]
0.01 µA
0.1 µA
0.001 mA
0.01 mA
0.0001 A
0.001 A
ACCURACY ± (% of reading + No. of Least Significant Digit)
FREQUENCY
30 Hz ~ 45 Hz
45 Hz ~ 1 kHz
1 kHz ~ 20 kHz
1.1+25
0.8+25
0.8+25
1.2+25
1.8+30[7]
0.9+25
0.9+30
0.9+25
0.9+25
3.3+70,
<3 A/5 kHz
U1251A/U1251B/U1252A/U1252B TEMPERATURE SPECIFICATIONS
THERMOCOUPLE
TYPE
K
J
(for U1252A)
RANGE
ºC
ºF
–200 ~ 1372 ºC
–328 ~ 2502 ºF
–210 ~ 1200 ºC
–346 ~ 2192 ºF
RESOLUTION
0.1 ºC/0.1 ºF
0.1 ºC/0.1 ºF
U1253A/U1253B TEMPERATURE SPECIFICATIONS
THERMOCOUPLE
TYPE
RANGE
ºC
ºF
–200 ~ –40 °C
–104 ~ –40 °F
K
–40 ~1372 °C
–40 ~ 2502 °F
–210 ~ –40 °C
–346 ~ –40 °F
J
–40 ~ 1200 °C
–40 ~ 2192 °F
RESOLUTION
0.1 ºC/0.1 ºF
0.1 °C /0.1 °F
0.1 °C /0.1 °F
0.1 °C /0.1 °F
CAPACITANCE SPECIFICATIONS
RANGE
RESOLUTION
10.000 nF
100.00 nF
1000.0 nF
10.000 µF
100.00 µF
1000.0 µF
10.000 mF
100.00 mF
0.001 nF
0.01 nF
0.1 nF
0.001 µF
0.01 µF
0.1 µF
0.001 mF
0.01 mF
ACCURACY
1% + 8
1% + 5
3% + 10
ºC
0.3% + 3 ºC
0.3% + 3 ºC
ACCURACY
ºF
0.3% + 6 ºF
0.3% + 6 ºF
ºC
1% + 3 °C
1% + 1 °C
1% + 3 °C
1% + 1 °C
ACCURACY
ºF
1% + 5.4 °F
1% + 1.8 °F
1% + 5.4 °F
1% + 1.8 °F
MEASURING RATE AT
FULL SCALE
4 times/sec.
1 time/sec.
0.1 times/sec.
0.01 times/sec
MAX. DISPLAY
11000 counts
5