English
Language : 

35670A Datasheet, PDF (7/10 Pages) Agilent(Hewlett-Packard) – Dynamic Signal Analyzer
Trigger
Trigger modes
Internal, source,
external (analog
setting) GPIB
Maximum trigger delay
Post trigger
8191 seconds
Pre trigger
8191 sample periods
No two channels can be further than
±7168 samples from each other.
External trigger max. input
±42 Vpk
External trigger range
Low range
-2 V to +2 V
High range
-10 V to +10 V
External trigger resolution
Low range
15.7 mV
High range
78 mV
Tachometer
Pulses per Revolution
RPM
RPM Accuracy
Tach level range
Low range
High range
Tach level resolution
Low range
High range
Maximum tach input level
Minimum tach pulse width
Maximum tach pulse rate
0.5 to 2048
5 ≤ RPM ≤ 491,519
±100 ppm (0.01%)
(typical)
-4 V to +4 V
-20 V to +20 V
39 mV
197 mV
±42 Vpk
600 nSec
400 kHz (typical)
Source output
Source types
Sine, random noise,
chrip, pink noise,
burst random, burst
chirp
Amplitude range
AC: ±5 V peak*
DC: ±10 V*
* Vacpk + |Vdc| ≤ 10 V
AC amplitude resolution
Voltage > 0.2 Vrms
2.5 mVpeak
Voltage < 0.2 Vrms
0.25 mVpeak
DC offset accuracy
±15 mV ± 3% of
(|DC| + Vacpk) set-
tings
Pink noise adder
Add 600 mV typical
when using pink
noise
Output impedance
<5Ω
Maximum loading
Current
±20 mA peak
Capacitance
0.01 µF
Sine amplitude accuracy at 1 kHz ±4% (0.34 dB) of
setting
Rload > 250 Ω
0.1 Vpk to 5 Vpk
Sine Flatness (relative to 1 kHz) ±1 dB
0.1 V to 5 V peak
Harmonic and sub-harmonic distortion and spurious
signals (In band)
0.1 Vpk to 5 Vpk sine wave
Fundamental < 30 kHz
< -60 dBc
Fundamental > 30 kHz
< -40 dBc
Digital interfaces
External keyboard
Compatible with
PC-style 101-key
keyboard
GPIB
Conforms to the following standards:
IEEE 488.1 (SH1, AH1, T6, TE0, L4, LE0, SR1, RL1,
PP0, DC1, DT1, C1, C2, C3, C12, E2)
EEE 488.2-1987
Complies with SCPI 1992
Data transfer rate
< 45 mSec for a
(REAL 64 Format)
401 point trace
Serial port
Parallel port
External VGA port
7