English
Language : 

U1230 Datasheet, PDF (6/10 Pages) Agilent(Hewlett-Packard) – Handheld Digital Multimeter (DMM)
Electrical Specifications
Frequency sensitivity specifications
For voltage measurements
Input range
Maximum input for specified accuracy1
600 mV in Scale mode
600 mV
6V
60 V
600 V
Minimum sensitivity (rms sine wave)
5 Hz to 50 kHz
U1231A
U1232A
U1233A
50 mV
50 mV
50 mV
120 mV
120 mV
120 mV
0.6 V
0.6 V
0.6 V
5.0 V
5.0 V
5.0 V
50 V
50 V
50 V
Notes:
1. Maximum input for specified accuracy, refer to “AC specifications” on page 106 of the User Guide.
For current measurements
Input range
Maximum input for specified accuracy1
60 μA
600 μA
6A
10 A
Minimum sensitivity (rms sine wave)
45 Hz to 5 kHz
U1232A
U1233A
30 μA
30 μA
30 μA
30 μA
0.5 A
0.5 A
0.5 A
0.5 A
Notes:
1. Maximum input for specified accuracy, refer to “AC specifications” on page 106 of the User Guide.
Scale transfer (mV)
Accuracy ± (% of reading + counts of least significant digit)
Range
Resolution
U1231A/U1232A/U1233A
DC 600 mV
0.1 mV
0.5% + 22
AC 600 mV
0.1 mV
1.0 % + 3 @ 45 Hz to 500 Hz
2.0 % + 3 @ 500 Hz to 1 kHz
Notes:
1. Overload protection: 600 Vrms for short circuits with < 0.3 A current.
2. The accuracy of the DC 600 mV range is specified after the Null function is used to subtract the thermal effect (by shorting the test leads).
3. Input impedance: 10 MΩ (typical).
6