E6832A Datasheet, PDF (5/6 Pages) Agilent(Hewlett-Packard) – W-CDMA Calibration Application | |||
|
◁ |
Fast Device Tune Measurement
Allows simultaneous calibration of a deviceâs transmitter output
power and receiver input level across level and frequency in a
single sweep (per frequency band). The device must operate in
a test mode, which forces it to transmit a predeï¬ned series of
power steps at various uplink frequencies, and also forces it to
simultaneously tune its receiver to perform measurements (such
as RSSI) of the test setâs signal at various downlink frequencies
and power levels.
Input level range
Average Tx power
â61 to +28 dBm/3.84 MHz
Measurement interval
1 timeslot (667 µs)
Measurement ï¬lter
W-CDMA mean power
Power measurement accuracy¹
0 to â45 dB of expected power and 20 to 55 °C
â54 to +28 dBm/3.84 MHz < ±0.75 dB, typically < ±0.4 dB
â61 to < â54 dBm/3.84 MHz < ±0.85 dB, typically < ±0.5 dB
0 to â45 dB of expected power and 20 to 55 °C with 48-hours
warm-up before self alignment initiated
â54 to +28 dBm/3.84 MHz < ±0.85 dB, typically < ±0.4 dB
â61 to < â54 dBm/3.84 MHz < ±0.95 dB, typically < ±0.5 dB
Output level ranges (Rx level at)
RF IN/OUT port
â108 to â15 dBm/3.84 MHz
RF OUT ONLY port
â108 to â5 dBm/3.84 MHz
Composite signal absolute output level accuracy
(< ±10 °C and < 24 hours from last self alignment)
â108 to â15 dBm/
< ±1.0 dB, typically < ±0.65 dBm
3.84 MHz at RF IN/OUT port
â108 to â5 dBm/
< ±1.0 dB, typically < ±0.65 dBm
3.84 MHz at RF OUT ONLY port
Output level settling (Rx level within 0.1 dB)
Amplitude switch
< 1 ms, typically < 250 µs
Frequency switch within band < 1 ms, typically < 500 µs
Step size range
5 to 20 ms
Trigger setup
Sources
Immediate, rise
1. Additional accuracy error when using RF OUT ONLY port is < ±0.1 dB.