E6832A Datasheet, PDF (5/6 Pages) Agilent(Hewlett-Packard) – W-CDMA Calibration Application
Fast Device Tune Measurement
Allows simultaneous calibration of a device’s transmitter output
power and receiver input level across level and frequency in a
single sweep (per frequency band). The device must operate in
a test mode, which forces it to transmit a predefined series of
power steps at various uplink frequencies, and also forces it to
simultaneously tune its receiver to perform measurements (such
as RSSI) of the test set’s signal at various downlink frequencies
and power levels.
Input level range
Average Tx power
–61 to +28 dBm/3.84 MHz
Measurement interval
1 timeslot (667 µs)
Measurement filter
W-CDMA mean power
Power measurement accuracy¹
0 to –45 dB of expected power and 20 to 55 °C
–54 to +28 dBm/3.84 MHz < ±0.75 dB, typically < ±0.4 dB
–61 to < –54 dBm/3.84 MHz < ±0.85 dB, typically < ±0.5 dB
0 to –45 dB of expected power and 20 to 55 °C with 48-hours
warm-up before self alignment initiated
–54 to +28 dBm/3.84 MHz < ±0.85 dB, typically < ±0.4 dB
–61 to < –54 dBm/3.84 MHz < ±0.95 dB, typically < ±0.5 dB
Output level ranges (Rx level at)
RF IN/OUT port
–108 to –15 dBm/3.84 MHz
RF OUT ONLY port
–108 to –5 dBm/3.84 MHz
Composite signal absolute output level accuracy
(< ±10 °C and < 24 hours from last self alignment)
–108 to –15 dBm/
< ±1.0 dB, typically < ±0.65 dBm
3.84 MHz at RF IN/OUT port
–108 to –5 dBm/
< ±1.0 dB, typically < ±0.65 dBm
3.84 MHz at RF OUT ONLY port
Output level settling (Rx level within 0.1 dB)
Amplitude switch
< 1 ms, typically < 250 µs
Frequency switch within band < 1 ms, typically < 500 µs
Step size range
5 to 20 ms
Trigger setup
Sources
Immediate, rise
1. Additional accuracy error when using RF OUT ONLY port is < ±0.1 dB.