E6832A Datasheet, PDF (3/6 Pages) Agilent(Hewlett-Packard) – W-CDMA Calibration Application | |||
|
◁ |
DPCH Measurement Suite
Includes mean and RRC-ï¬ltered mean power, spectrum emission
mask, ACLR, composite EVM frequency error, PCDE, and occupied
bandwidth measurements.
Trigger setup
Delay
Sources
â20 to +20 ms
External, fall, immediate, rise
Mean and RRC-ï¬ltered mean power
measurements
Input level range
Average power
â61 to +28 dBm/3.84 MHz
Measurement accuracy¹
â54 to +28 dBm/3.84 MHz < ±0.6 dB, typically < ±0.3 dB
â61 to < â54 dBm/3.84 MHz < ±0.7 dB, typically < ±0.4 dB
With < 48 hours warm-up before self alignment initiated
â54 to +28 dBm/3.84 MHz < ±0.7 dB, typically < ±0.3 dB
â61 to < â54 dBm/3.84 MHz < ±0.8 dB, typically < ±0.4 dB
Measurement repeatability Typically < ±0.05 dB
Filter
Mean power
RRC-ï¬ltered mean power
Mean power-compliant ï¬lter
3.84 MHz
Spectrum emission mask (SEM) measurement
This measurement provides a ratio of the transmitted power
in a 3.84 MHz RRC bandwidth to offset frequencies, which
are between 2.5 MHz and 12.5 MHz away from the UE carrier
frequency.
Input level range
Average power
+5 to +28 dBm/3.84 MHz
Measurement accuracy
2.5 to 12.5 MHz
< ±1.5 dB, typically < ±0.5 dB
Offset frequency measurement bandwidths
2.5 to 3.5 MHz offsets
30 kHz
3.5 to 12.5 MHz offsets
1 MHz
In-band measurement ï¬lter 3.84 MHz (RRC)
Adjacent channel leakage ratio (ACLR)
measurement
This measurement provides a ratio of the ï¬ltered mean trans-
mitted power to the ï¬ltered mean power in an adjacent channel.
Both the transmitted and the adjacent channel powers are
measured with a ï¬lter that has an RRC response with roll-off a =
0.22 and a bandwidth equal to the chip rate.
Input level range
Average power
Measurement accuracy
±5 MHz offsets at â33 dB
±10 MHz offsets at â43 dB
Residual noise ï¬oor
±5 MHz offsets
±10 MHz offsets
Filter
+5 to +28 dBm/3.84 MHz
< ±0.8 dB, typically < ±0.4 dB
< ±0.8 dB, typically < ±0.4 dB
Typically < â53 dBc
Typically < â63 dBc
3.84 MHz (RRC)
Error vector magnitude (EVM) measurement
This measurement calculates composite EVM and several other
results relating to UE modulation quality.
Input level range
Average power
â25 to +28 dBm/3.84 MHz
UE ranges
EVM
Frequency error
Timing error
⤠35% rms
< ±1 kHz
< ±50 chips
EVM measurement accuracy (Includes residual EVM)
UE EVM 0% rms
< 2.5% rms
UE EVM 17.5% rms
< 0.7% rms
UE EVM 35% rms
< 0.5% rms
Filter
3.84 MHz (RRC)
Measurement results
EVM, magnitude error, phase
error, origin offset
1. Additional accuracy error when using RF OUT ONLY port is < ±0.1 dB.