E7400A Datasheet, PDF (3/8 Pages) Agilent(Hewlett-Packard) – EMC Analyzers | |||
|
◁ |
Amplitude specifications
Amplitude range
Measurement range
Displayed average noise level
(DANL) to maximum safe input
level
Input attenuator range
E7402A
0 to 65 dB (75 dB1), in 5 dB steps
E7405A
0 to 65 dB, in 5 dB steps
Maximum safe input level
Average continuous power +30 dBm (1 W)
Peak pulse power
+50 dBm (100 W)
(input attenuator ⥠30 dB)
Maximum dc
0 Vdc (dc coupled)
50 V (ac coupled)
1 dB gain compression (total power at input mixer2)
⥠50 MHz
0 dB
⥠6.7 GHz
â3 dB
⥠13.2 GHz
â5 dB
Display range
Log Scale
0.1, 0.2, 0.5 dB/division and 1 to 20 dB/
division in 1 dB steps; ten divisions
displayed
RBW ⥠1kHz 0 to â85 dB from reference level is calibrated
RBW ⤠300 Hz 0 to â1205 dB from reference level is calibrated
Linear scale 10 divisions
Scale units dBm, dBmV, dBµV, dBµA, Amps, Volts and
Watts
Marker readout resolution
Log scale
0 to â85 dB
0.04 dB
0 to â120 (RBW ⤠300 Hz) 0.04 dB
Linear scale
0.01% of reference level
Fast sweep times for zero span (Option AYX)
Log Scale, 0 to â85 dB
0.3 dB
Linear
0.3 % of reference level
Display average noise level (DANL)
1 kHz
RBW
E7402A
30 Hz to 9 kHz3
9 kHz to 100kHz3
100 kHz to 1 MHz3
1 MHz to 10 MHz3
10 MHz to 1 GHz
1 GHz to 2 GHz
2 GHz to 3 GHz
E7405A
30 Hz to 9 kHz3
9 kHz to 100kHz3
100 kHz to 1 MHz3
1 MHz to 10 MHz3
10 MHz to 1 GHz
1 GHz to 2 GHz
2 GHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 12 GHz
12 GHz to 22 GHz
22 GHz to 26.5 GHz
na
na
na
⤠â117
⤠â117
⤠â116
⤠â114
na
na
na
⤠â117
⤠â116
⤠â116
⤠â112
⤠â112
⤠â111
⤠â107
⤠â106
10 Hz
RBW
⤠â93
⤠â109
⤠â135
⤠â136
⤠â136
⤠â135
⤠â133
⤠â93
⤠â109
⤠â135
⤠â137
⤠â135
⤠â131
⤠â131
⤠â131
⤠â130
⤠â126
⤠â125
1 kHz
w/preamp
on
na
na
na
na
⤠â1524
⤠â1534
⤠â1514
na
na
na
na
⤠â1514
⤠â1514
⤠â1494
na
na
na
na
10 Hz
w/preamp
on, typical
na
na
na
⤠â152
⤠â156
⤠â156
⤠â154
na
na
na
⤠â155
⤠â157
⤠â155
⤠â152
na
na
na
na
1 Hz
w/preamp
on, typical
na
na
na
⤠â162
⤠â166
⤠â166
⤠â164
na
na
na
⤠â165
⤠â167
⤠â165
⤠â162
na
na
na
na
1. Characteristic
2. Mixer power level (dBm) = input power (dBm) â input attenuator (dB)
3. Typical
4. 0 to 50 dB for RBWs ⤠300 Hz and span = 0 Hz, or when auto ranging is off,
or 0 to 30 dB for RBW = 200 Hz.
5. 0 to â70 dB range when span = 0 Hz, when RBW = 200 Hz, or when IF gain
is fixed.
3
|
▷ |