English
Language : 

E7400A Datasheet, PDF (1/8 Pages) Agilent(Hewlett-Packard) – EMC Analyzers
Agilent
E7400A Series EMC Analyzers
Data Sheet
These specifications apply to the Agilent
Technologies E7402A and E7405A EMC
analyzers.
Frequency Specifications
Frequency range
E7402A
dc coupled
100 Hz1 to 3.0 GHz
ac coupled
100 kHz1 to 3.0 GHz
E7405A
Band LO harmonic = N
0 1- dc coupled 30 Hz1 to 3.6 GHz
ac coupled 100 MHz to 3.6 GHz
1 1-
2.85 GHz to 6.7 GHz
2 2-
6.2 GHz to 13.2 GHz
3 4-
12.8 GHz to 19.2 GHz
4 4-
18.7 GHz to 26.5 GHz
Specifications
All specifications apply over 0 °C to +55 °C unless other-
wise noted and are covered by the product warranty. The
analyzer will meet its specifications when: it’s within the
one year calibration cycle, AUTO ALIGN [ALL] is selected,
stored a minimum 2 hours within the operating tempera-
ture range, turned on for at least 5 minutes, and Align
Now RF has been run once every 24 hour period. Typical per-
formance describes the level at which 80% of the units will
meet or exceed with a 95% confidence level over 20 to 30 °C,
but is not covered in the product warranty. Characteristics
describe expected product performance levels that are not
covered in the product warranty.
Frequency reference
Aging
Temperature stability
Settability
±1 x 10 –7/year
±1 x 10 –8
±1 x 10 –8
Frequency readout accuracy
(start, stop, center, marker) ±(frequency indication
x frequency reference error2
+ span accuracy + 15% of RBW
+ 10 Hz) + 1 Hz x N3
1. Usable to 30 Hz
2. Frequency reference error = (aging rate x period of time since
adjustment + settability + temperature stability)
3. N = LO harmonic mixing mode