English
Language : 

4395A Datasheet, PDF (1/20 Pages) Agilent(Hewlett-Packard) – Network/Spectrum/Impedance Analyzer
Agilent
4395A
Network/Spectrum/Impedance Analyzer
Data Sheet
Network Measurement
1. At relative to 0 dBm output, 50 MHz,
23 °C ± 5 °C
Specifications describe the instrument’s warranted performance over
the temperature range of 0 °C to 40 °C (except as noted). Supplemental
characteristics are intended to provide information that is useful in applying the instru-
ment by giving non-warranted performance parameters. These are denoted as SPC
(supplemental performance characteristics), typical, or nominal. Warm up time must be
greater than or equal to 30 minutes after power on for all specifications.
Source characteristics
Frequency characteristics (Option 4395A-800)
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Hz to 500 MHz
Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 mHz
Frequency reference
Accuracy
at 23 °C ± 5 °C, referenced to 23 °C . . . . . . . . . . . . . . . . . . . . . . . . . . < ±5.5 ppm
Aging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < ±2.5 ppm/year (SPC)
Initial achievable accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . < ± 1.0 ppm (SPC)
Temperature stability
at 23 °C ± 5 °C, referenced to 23 °C . . . . . . . . . . . . . . . . . . . . . < ±2 ppm (SPC)
Precision frequency reference (Option 4395A-1D5)
Accuracy
at 0 °C to 40 °C, referenced to 23 °C . . . . . . . . . . . . . . . . . . . . . . . . < ±0.13 ppm
Aging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < ±0.1 ppm/year (SPC)
Initial achievable accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . < ±0.02 ppm (SPC)
Temperature stability
at 0 °C to 40 °C, referenced to 23 °C . . . . . . . . . . . . . . . . . . . < ±0.01 ppm (SPC)
Output characteristics
Power range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –50 dBm to + 15 dBm
Level accuracy
at 0 dBm output, 50 MHz, 23 °C ± 5 °C, . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1.0 dB
Level linearity
Output power
≥ -40 dBm
< -40 dBm
Linearity1
±1.0 dB
±1.5 dB