Precautions (electrostatic countermeasures)
・ When handling CCD sensors, always wear a wrist strap and also anti-static clothing, gloves, and shoes, etc. The wrist strap should
have a protective resistor (about 1 MΩ) on the side closer to the body and be grounded properly. Using a wrist strap having no
protective resistor is hazardous because you may receive an electrical shock if electric leakage occurs.
・Avoid directly placing these sensors on a work bench that may carry an electrostatic charge.
・Provide ground lines with the work bench and work floor to allow static electricity to discharge.
・Ground the tools used to handle these sensors, such as tweezers and soldering irons.
It is not always necessary to provide all the electrostatic measures stated above. Implement these measures according to the
amount of damage that occurs.
Element cooling/heating temperature incline rate
When cooling the CCD by an externally attached cooler, set the cooler operation so that the temperature gradient (rate of tempera-
ture change) for cooling or allowing the CCD to warm back is less than 5 K/minute.
Related information
www.hamamatsu.com/sp/ssd/doc_en.html
Precautions
∙ Notice
∙ Image sensors/Precautions
Technical information
∙ FFT-CCD area image sensor/Technical information
Features
Multichannel detector head C7040
Area scanning or line-binnng operation
Readout frequency: 250 kHz
Readout noise: 20 e- rms
Input
Supply voltage
Master start
Master clock
Symbol
VD1
VA1+
VA1-
VA2
φms
φmc
Specification
+5 Vdc, 200 mA
+15 Vdc, +100 mA
-15 Vdc, -100 mA
+24 Vdc, 30 mA
HCMOS logic compatible
HCMOS logic compatible,
1 MHz