English
Language : 

C9730DK-10_11 Datasheet, PDF (6/8 Pages) Hamamatsu Corporation – For soft X-ray imaging, cassette type with USB 2.0 interface Photodiode area: 52.8 × 52.8 mm
Flat panel sensor
C9730DK-10
Dimensional outline (unit: mm, tolerance: ±1 mm unless otherwise noted)
Screw hole *
Screw hole *
18.5 31 23.5 12.5 25 12.5
14.5
CMOS FLAT PANEL SENSOR C9730DK-10
Active area
51.6
(1031, 1031)
(0, 0)
108
266
C1.3
Scintillator to
top cover
5.4
Top cover material including window is carbon fiber (1.0 mm thickness).
Weight: 1.3 kg
* M3 screw hole for fixing USB plug to flat panel sensor.
The depth of the hole: 7 mm
261 ± 0.5
(6 ×) M3 depth 8
151 ± 0.5
57.5 ± 0.5
KACCA0211ED
KACCA0211ED
Notice
· Do not subject the flat panel sensors to strong vibration or shock (Strong shock such as drop impacts may cause permanent damage
to these sensors)
· Users must take responsibility for implementing X-ray shielding safety measures to avoid the risk of X-ray exposure.
· Data listed in this datasheet is defined at the time of shipment. Characteristics may vary somewhat due to exposure to X-rays so take
proper countermeasures such as making periodic image correction.
· This product is warranted for a period of 12 months after the date of the shipment.
The warranty is limited to replacement or repair of any defective product due to defects in workmanship or materials used in manufacture.
The warranty does not cover loss or damage caused by natural disaster, misuse (including modifications and any use not complying
with the environment, application, usage and storage conditions described in this datasheet), or total X-ray radiation dose over 45000
Roentgen [incident X-ray energy: less than 35 kVp (17 keV Mo source)] even within the warranty period.
The C9730DK-10 conforms to European EMC directives:
EN61326-1 Class A.
6