English
Language : 

C9730DK-10_11 Datasheet, PDF (3/8 Pages) Hamamatsu Corporation – For soft X-ray imaging, cassette type with USB 2.0 interface Photodiode area: 52.8 × 52.8 mm
Flat panel sensor
C9730DK-10
Resolution
100
90
80
70
60
50
40
30
20
10
0
01
(Ta=25 °C, A.vdd=5.0 V, D.vdd=5.0 V)
2345678
Spatial frequency (line pairs/mm)
9 10
KACCB0154EB
Accessories
O Power cable (terminated with a JSTJ-C9-2C plug at one end and open at the other end; 2 m; see Table 1.)
O External trigger cable (terminated with an FGG.0B.304.CLAD56 plug at one end and open at the other end; 5 m; see Table 2.)
O Earth cable (AWG 18; 4 m)
System requirements
To operate the C9730DK-10 at full performance, the following system and peripherals are required.
O PC: IBM compatible PC with USB 2.0 running on Windows XP
O Power source: A.vdd = +5.0 ± 0.1 V (800 mA), D.vdd = +5.0 ± 0.1 V (300 mA)
· The voltages described above are specified at the flat panel sensor side.
· Please use a low noise series power supply. (Avoid using a switching power supply.)
· Install a noise filter on the AC power input line to prevent surges on the AC line.
· Always ground the ground terminal to avoid the effects of noise from peripheral devices.
O DCAM-API (digital camera application programming interface): produced by HAMAMATSU
· The driver software, DLL and sample software are included in DCAM-API.
· The latest version of DCAM-API can be downloaded from “http://www.dcamapi.com”.
O USB cable:mini-B type